133
FLIM Data Acquisition
16.1.2.5
Using White Light Lasers
1. Select the desired laser lines in the
Whitelight
dialog in
Beam Path Settings
.
2. Make sure that the AOBS is at fluorescence.
3. Set the desired intensity using the AOTF slider.
4. If the system has a multifunction port (
MFP
), it should be set to
Substrate
.
5. Open the
Acquisition
tab and adjust the rotation of the scan field to 0° using the slider
).
Figure 118:
Adjusting the diameter of the pinhole and rotation of the scan field
16.1.3
Adjusting the Pinhole for Image Acquisition
For VIS and UV excitation, set the
Pinhole
diameter to
1 Airy
; for MP excitation set it to the
maximum size (
).
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......