CIAN Leica SP8 – short instructions, version 3.1, September 2015 

page 3 of 10

 

 

2.2 - Cleaning objectives

 

Clean immersion objectives before and after use. The dry objectives on the SP8 should not be 
cleaned in this fashion since they have concave front lenses. 
Use 

dry lens paper 

to blot off larger amounts of oil or other immersion medium: 

  Take a sheet of lens paper, fold it in three into a long rectangle; do not run your fingers 

over the middle of the paper that will touch the lens 

  Wipe the lens by holding either end of the lens paper, dragging it gently across the 

objective lens three times. Use a fresh area of the tissue each time. 

  Repeat as needed with a fresh lens paper to remove excess of immersion oil. 

Use 

wet lens paper

 to remove traces of oil, other dirt: 

  Fold as above, put a drop or two of lens cleaner (the blue fluid) on the lens paper. 

  Wipe the lens gently by holding either end of the wet lens tissue dragging it gently across 

the objective lens three times. Use a fresh area of the tissue each time. 

  Repeat with water. 

NEVER wipe the lens in a circular pattern; NEVER apply any pressure directly to the lens. 

Note: 

If you find something on the lens that you can’t remove with this procedure,

 please 

contact CIAN staff as soon as possible, and discontinue use of the objective 

until the 

problem is solved 

 
 
 
 
 

3 - Operating Leica SP8  

3.1 - Manual controls 

Note that you can use the TL and WFF functions independent of the computer and the 
scanner/laser power. 

 

1: WFF filter cubes (blue, green, red) 
2: WFF shutter 
3: status display 
 
4: XYZ joystick (

SmartMove

) with separate 

coarse/fine toggle buttons for x/y, z 
 
5: TL activation, toggle BF, DIC, Pol 
6: focus knob 
7: Field aperture (AP) control 
8: BF light intensity 
 
9: move objectives up/down; set Z 
zero/limits (hold ‘set’, push up arrow twice) 
10: XYZ coarse/fine toggle 

 
 
 
 

Fig  4

: Microscope front and side panels 

Summary of Contents for SP8 AT CIAN

Page 1: ...oscope 1 Leica DMI 6000B inverted microscope 2 Scan head 3 Anti vibration table 4 EL6000 metal halide lamp for WFF 5 CTR control box for microscope stand 6 Smart Move joystick 7 Monitor 8 Control pane...

Page 2: ...t the condenser or objectives Initialization is needed for Tiling and Mark Find features 7 Visually inspect and clean objectives see section 2 2 as needed 2 1 Adjusting K hler Illumination 1 Insert a...

Page 3: ...ue fluid on the lens paper Wipe the lens gently by holding either end of the wet lens tissue dragging it gently across the objective lens three times Use a fresh area of the tissue each time Repeat wi...

Page 4: ...mode by clicking the Laser icon Activate the lasers you need by clicking on their on off button The standby feature is not required 3 2 3 Useful settings in the configuration tab Customize the USB Co...

Page 5: ...ion 1 Click SEQ in the Acquisition Mode to open the Sequential Scan control window 2 Set light path configuration for the first sequence as described above in 3 2 4 a 3 Click in the Sequential Scan co...

Page 6: ...space Rotation of the scan field by up to 100 degrees in either direction as with zoom use slider numerical factor or control panel knob Panning with the arrows positions the scan area when zoomed in...

Page 7: ...positioning during the stack Note Z Wide refers to z control via the objective focus drive 4 Zoom into cartoon view to visualize schematic stack representation 5 Z step size or number of steps can be...

Page 8: ...ect it in the Positions list 5 Using the buttons marked positions can be a Saved and reloaded b individually deleted c all deleted trash can d edited go to position using positions list adjust xyz pre...

Page 9: ...orage You are responsible for safe and redundant storage of your data on your own storage devices 1 Save projects in LAS X software to your folder on the D drive 2 Connect to the server Run command co...

Page 10: ...1 water HC PL APO 40x 1 10 W CORR CS2 DIC 5 40x 1 3 oil HC PL APO 40x 1 30 Oil CS2 DIC 6 63x 1 4 oil HC PL APO 63x 1 40 OIL CS2 DIC Confocal mode Lasers no fluorescence emission filters all spectral...

Reviews: