Operation
34
PROVIDO Ref. 10732418 / Version 03
PROVIDO Ref. 10732418 / Version 03 35
8.3.5 Adjusting the magnification (zoom)
You can adjust the magnification using a footswitch/handle or the
"Magnification" adjustment bar on the "Main" menu screen of
the control unit.
On the touch panel monitor in the "Main" menu screen:
X
Press the the
or
button on the bar for adjusting the
magnification.
– or –
X
Press the magnification adjustment bar directly.
The magnification changes.
• Clicking the
or
button changes the
magnification value in increments of 1. Holding down
the button with your finger changes the value in
increments of 5.
• You can adjust the magnification motor speed in the
"Speed" menu .
• These values can be saved individually for each user (see
page 41).
WARNING
Danger to the patient due to failure of the magnification
motor.
X
If the magnification motor fails, adjust the
magnification manually.
You can read the currently set magnification on the
display (1) on the M530 optics carrier.
1
Manually adjusting the magnification (zoom)
CAUTION
Destruction of the magnification motor.
X
Only adjust the magnification manually if the
magnification motor is defective.
If the magnification motor fails, the magnification can be manually
adjusted using the rotary knob (2).
X
Push in rotary button (2).
X
Set the desired magnification by turning the knob.
2