background image

19

Leica MATS – User Manual

Notes

Summary of Contents for MATS

Page 1: ...Leica MATS User manual...

Page 2: ...ontents Contents 2 Page Description 3 Safety concept 4 Assembly The components 8 Installation 10 Use Operating controls 12 Switching on and off 13 Temperature control 13 Dimensions and weight 14 Techn...

Page 3: ...operation of the Leica thermocontrol system MATS Your instrument is supplied with a printed user manual in English Additional languages and information can be found on the inter active CD ROM that is...

Page 4: ...a control unit a power cable and various thermoplate stage types for Leica stereomicroscopes transmitted light stands and micro scopes upright inverted Thermoplate and control unit are matched to each...

Page 5: ...able surface If the specimen in a petri dish or on a slide glass indicates a lower temperature compared to the displayed process value adjust the set value to obtain a proper specimen temperature Othe...

Page 6: ...jury through electric shock Only use the power cables listed on page 16 Inspect the power cable regularly for damage If it is defec tive replace it immediately because it could cause the instrument an...

Page 7: ...unit Respect the warning signs relating to the handling of liquids Do not soak into water or into solvent Never use any solvents such as gasoline or thinner to wipe the instrument The color of the sur...

Page 8: ...ith transmitted light base lightfield darkfield 4 Thermoplate type MATS URS LT for inverted microscopes Leica DMIRB 3 plate x y stage 5 Thermoplate type MATS UDM LT for inverted microscopes Leica DMIR...

Page 9: ...9 Leica MATS Assembly MATS MZ367 LT MATS LZ LT MATS URS LT MATS UDM LT MATS U55S LT MATS STD LT...

Page 10: ...6 Heating stage type MATS MZ367 LT for Leica M stereomicroscopes with trans mitted light base HL Tighten attachment screws on both sides of the thermoplate Heating stage type MATS LZ LT for Leica M st...

Page 11: ...RB DMIL Heating stage type MATS U55S LT for upright microscopes general Attach heating stage with adapter according to the figure Heating stage type MATS STD LT standard Connect the thermoplate with t...

Page 12: ...Temperature control 3 Setting the temperaturs Up Down keys 4 PV display shows the process value and various symbols 5 SV display shows different settings values Operating controls 12 Leica MATS Operat...

Page 13: ...oper specimen temperature Otherwise the spec imen could be damaged With stereomicroscopes it must be ensured that the existing glass insert is removed from the base before the heating stage is placed...

Page 14: ...with transmitted light base lightfield darkfield Weight about 540 g Thermoplate type MATS URS LT for inverted microscopes Leica DMIRB 3 plate x y stage Weight about 100 g Thermoplate type MATS UDM LT...

Page 15: ...s and weight 202 174 257 265 10 MATS MZ367 LT 145 113 165 61 100 80 6 116 80 142 115 9 2 5 205 180 7 160 185 MATS LZ LT MATS STD LT MATS U55S LT MATS URS LT MATS UDM LT 146 220 196 332 10 3 132 150 13...

Page 16: ...cord set 3 conductor grounding type SVT No 18 AWG rated at 125 V 7 A minimum if an extension cord is used use only a power supply cable with PE protective earth connector For mains supplies with 220...

Page 17: ...lue t 37 C step 0 1 C setting method with Up and Down key setting range room temperature to 50 C achievable temperature accuracy 0 3 C with indicator temperature sensor thermocouple connection to heat...

Page 18: ...18 Leica MATS User Manual Notes...

Page 19: ...19 Leica MATS User Manual Notes...

Page 20: ...e solutions to our customers needs for vision measurement lithography and analysis of microstructures Leica the leading brand for microscopes and scientific instruments has developed from five brand n...

Reviews: