Adjusting
1. Set the dioptric correction for both
eyepieces to "0".
2. Select the lowest magnification and focus
on a flat specimen.
3. Select the highest magnification and read-
just the sharpness.
4. Select the lowest magnification again, but
do not look into the eyepieces.
6. Rotate the eyepieces counterclockwise in
the "+" direction as far as they will go (+5
diopter settings).
Leica stereomicroscopes are parfocally
matched. The prerequisite for this is the correct
setting of the diopters and the parfocality. The
following adjustments have to be carried out
only once by each user.
Preparations
★
Move the lever of the video/phototube to
the "observation" position and open the
diaphragm.
7. Look into the eyepieces.
8. Slowly rotate each eyepiece individually
in the "–" direction until each eye sees the
object sharply imaged.
9. Select the highest magnification and refo-
cus if necessary.
Now, if you adjust the magnification from the
lowest to the highest level, the specimen is
always brought into sharp focus. If not, repeat
the process.
Dioptric Correction and Parfocality
Summary of Contents for M60 B
Page 1: ...Living up to Life Leica M60 B Leica M80 B User Manual...
Page 13: ...Leica M series User Manual 13 Leica M Series...
Page 17: ...Leica M series User Manual 17 Assembly...
Page 25: ...Leica M series User Manual 25 Quick Start Guide...
Page 34: ...Leica M series User Manual 34 Eyepieces...
Page 40: ...Leica M series User Manual 40 Objectives and Optical Accessories...
Page 42: ...Leica M series User Manual 42 Bases...
Page 49: ...Leica M series User Manual 49 Specifications for the Bases...