60
Version 2.0, Revision I
Cleaning and Maintenance
6
The printer is once again ready for printing.
Warning
If no button is pressed to acknowledge the end of the cleaning procedure, the print head will be
closed automatically after a few minutes to prevent it from drying out.
However, the message
15
remains on the display, as the instrument assumes that the cleaning has
not been carried out.
6.3
Exchanging the cartridge
Note
The ink cartridge has to be replaced after 3.5 months at the latest or after 60,000 prints. On the white
surface on the front side of the ink cartridge, note the date when the ink cartridge was installed.
6.3.1 Removing the used ink cartridge
on the left side of the instrument by pressing its top left
corner
and loosen it again with one full turn.
3. Press the red retaining bracket
down and pull the ink cartridge out
approx. 30 mm until the
INK EMPTY
LED
4. Retighten the red cap and completely remove the cartridge from the printer.
5. Store the ink cartridge in a horizontal position in a sealed container.
6. Dispose the used ink cartridge in accordance with the regulations of your laboratory and your
legislative.
Fig. 54
1
2
3
4