57
Leica IP S
Cleaning and Maintenance
6
1
2
3
Fig. 49
•
The drawer for broken glass
is locatedabove the slot
for the ink cartridge, behind the left cover
. It prevents glass dust and
broken glass from falling into the interior
of the instrument. The drawer can be
pulled out sideways by grasping the black
handle
regularly.
•
Broken glass can easily be removed - use
a brush and remove any glass through the
opening in the center of the drawer.
Outer surfaces
•
Clean the outer surfaces (including those of the automated
slide
unload station) with a mild detergent
and subsequently dry with a slightly moistened cloth.
•
Do not use any solvents for cleaning the outer surfaces and the lid!
Automated unload station
•
Remove the trays; with a brush, remove dust and debris from guides and ejector.
•
The trays themselves can be cleaned with a household cleaner.
•
Do not use any solvents to clean the trays!
•
Prior to reinserting them into the instrument, the trays must be completely dry.
6.2
Cleaning the print head
Preparing the printer:
The print head must be cleaned manually once a week or if the message
15
is displayed.
1. Open the printer hood
and then press the
CLEAN
and
LOADED
keys
simultaneously.
moves upward to a position approx. 1 cm away from the sealing