Embedding mold tray and cassette tray
• Depending on the preferred working direction, the two heated instrument trays
can be used for embedding molds or cassettes. The temperature is adjustable between
50 °C (122 °F) and 75 °C (167 °F).
for cassettes or embedding molds can be placed in each instrument
tray.
• A lid
is provided for each of the trays
to prevent loss of heat and
contamination inside the tray. For easy access, the lid can be folded up.
Warning
• Only use the cassette, embedding mold tray and paraffin tank with the supplied lids.
• Wear gloves when opening the lids.
• It is recommended to clean the cassette tray before adding new specimen.
Note
• When working with half-open lids
, adjust the temperature to ensure that the
paraffin remains molten.
• Use the removable trays in the cassette/mold tray in normal work-flow.
• Use the compatible basket with the trays. Use correct molds.
• Make sure all cassettes are fully covered during operation.
• Do not fill excessive paraffin to the cassette/mold tray. Risk of burning!
Fig. 13
9
10
15
• Press the power switch ON/OFF to connect/cut the main power.
"I" =
ON
"O" =
OFF
• A lamp lights up in the switch to indicate that the instrument is connecting to the main power.
• After the instrument has been commissioned, the power switch should only be used if the instrument
is to be switched off for a longer period of time.