Instructions for Use

Leica HI1220

Flattening Table

Leica HI1220 V 3.3 , English 07/2016

Order No.:

 14 0423 80101 RevG

Always keep this manual with the instrument.
Read carefully before working with the instrument.

Summary of Contents for HI1220

Page 1: ...ructions for Use Leica HI1220 Flattening Table Leica HI1220 V 3 3 English 07 2016 Order No 14 0423 80101 RevG Always keep this manual with the instrument Read carefully before working with the instrum...

Page 2: ......

Page 3: ...contents or technical details of the present Instructions for Use are not to be considered warranted characteristics of our products These are determined only by the contract provisions agreed between...

Page 4: ...3 1 Technical data 9 4 Instrument Setup 10 4 1 Installation site requirements 10 4 2 Standard delivery 10 4 3 Unpacking and setting up the HI1220 11 4 4 Setting up the power supply 11 5 Operation 12 5...

Page 5: ...appear in a gray box and are marked by a warning triangle Observe the instructions for use Manufacturer Notes i e important user information appear in a gray box and are marked by an information symb...

Page 6: ...type indicated on the cover page A nameplate indicating the instrument serial num ber is attached to the rear side of the instrument Any other use of the instrument is considered improper Failure to...

Page 7: ...caution notes in this chapter must be observed at all times Be sure to read these notes even if you are already familiar with the operation and use of other Leica products The protective devices locat...

Page 8: ...t must be switched off and unplugged from the power supply during all repair and service work During operation the instrument surface can be very hot Risk of injury Do not store any combustible and fl...

Page 9: ...ed For 100 120 V 5x20 mm 2x T 5A L250 VAC For 230 240 V 5x20 mm 2x T 2 5A L250 VAC Dimensions 350x310x100 mm Empty weight 3 6 kg w o packaging Relative humidity 20 80 non condensing IEC 1010 classific...

Page 10: ...ight and excessive temperature fluctuations The instrument must be connected to a suitable power socket Use only the provided power cable which is intended for the local power supply 4 2 Standard deli...

Page 11: ...ton Fig 4 Hold the instrument firmly with one hand pull off both foam sections Fig 5 and remove the dust cover 4 4 Setting up the power supply Before connecting the power cable make sure that the powe...

Page 12: ...n with the main power switch 6 SET Indicatestheselectedtargettempera ture 7 Set key lower temperature 8 Set key increase temperature 9 RUN STOP Switch instrument on or off 10 Display 5 2 Switching the...

Page 13: ...atically This stored target value will continue to be displayed for about 2 seconds then the current actual temperature display will appear again The target temperature can be checked by pressing the...

Page 14: ...aboratory detergents The instrument must be switched off and have cooled down for cleaning Do not use organic solvents e g alcohol or xylene 6 2 Replacing fuses Switch the unit off and unplug it Use o...

Page 15: ...Leica sales organization or the organization from which you have purchased the contractual product shall apply exclusively Service information If you require technical service or replacement parts ple...

Page 16: ...t proper decontamination documentation If you have any further questions please call your local Leica organization Leica Internal Use If applicable note corresponding Job and RAN RGA Number Job Sheet...

Page 17: ...henever possible please use the original transportation case box Yes No 3 2 Signature Date Name Position eMail Institute Department Address Phone Fax Important to avoid refusal of shipment Place one c...

Page 18: ...18 Instructions for Use V 3 3 RevG 07 2016 Notes...

Page 19: ......

Page 20: ...www LeicaBiosystems com Leica Biosystems Nussloch GmbH Heidelberger Stra e 17 19 D 69226 Nussloch Phone 49 6224 143 0 Fax 49 6224 143 268 Web www LeicaBiosystems com...