Preparation before surgery
Leica FL800 ULT / Ref. 10 744 763 / Version 06
9
4.2 Checking the Leica FL800 ULT
illumination
WARNING
Risk of infection due to insterile Leica FL800 ULT test
patterns.
X
Check the Leica FL800 ULT illumination in non-sterile
environment only.
X
Place the Leica FL800 test pattern under the microscope.
X
Focus the microscope at a low magnification.
The Leica FL800 test pattern will be visible in the microscope in
white-light mode.
Take care to ensure the precise parfocal setting of the
Leica surgical microscope. Follow the instructions on
parfocal setup.
X
Activate Leica FL800 mode with the handle.
A connected monitor will display the Leica FL800 image of
the ICG test pattern.
Leica FL800 test pattern in white-light mode
• Front side
Leica FL800
ICG Fluorescence Standard
Please read
the Leica
FL800 user
manual!
10 713 880-xx
!
high
intensity
1
1 ICG test pattern
• Back side
low
intensity
Store below 30°C
and subdued light
wwyy.wwyy-nn
Expiry dd.mm.yyyy
Leica FL800 test pattern in FL800 mode
• With high intensity, use the front side
1
• With low intensity, use the back side
+
1
1 ICG test pattern