20
Version 1.7, Revision K
Safety
2
2. To unlock the handwheel, turn the handwheel brake lever
back to its original
position
.
• Using the handwheel locking mechanism
on the top of the handwheel, the handwheel
can be braked in the 12 o'clock position.
1. To lock the handwheel, press the handwheel locking mechanism
continue to turn clockwise the handwheel slowly until it locks exactly in the 12 o'clock position.
2. To unlock the handwheel, turn the handwheel locking mechanism
inwards.
Tip
When using both brake systems at the same time, always move the handwheel brake lever
to position
first. Otherwise, it may be not be possible to release the handwheel
locking mechanism
2.3.2 Safety guard on the knife/blade holder
Each knife holder is equipped with a tightly mounted safety guard
makes it possible to cover completely the cutting edge in every knife or blade position.
Fig. 2
1
2
2
Knife holder N
of the knife holder
N can be easily positioned via two handles
. To cover the knife edge, push both
cover strips of the safety guard to the center.
Fig. 3
1
2
3
Two-in-one blade holder E for high and low
profile blades
The safety guard on the two-in-one blade holder
E consists of a red foldaway handle
.
To cover the cutting edge, fold the safety guard
handle upwards as illustrated in
.
Tip
The clamping levers on the two-in-one blade holder E are not interchangeable. The two clamping
levers
must remain in the position shown at all times, as otherwise isolated
malfunctions of the two-in-one blade holder E can occur. The clamping lever for the blade
is at the right side, the clamping lever for the lateral displacement