18
Version 6.9, Revision H
Safety
3
Microtome sledge
Warning
Unintentional touching of the knife or unintentional movement of the microtome sledge
Severe cutting injuries and / or severe crushing injuries
•
Move the microtome sledge to the position which is most appropriate for your purpose and that
ensures the best safety.
•
Prior to any work involving the knife or the microtome, or inside the cryochamber, activate the
Emergency stop
p. 18 – 3.3.1 Emergency stop switch)
•
For any work involving the knife or the microtome, or inside the cryochamber, wear the safety gloves
provided, see
p. 23 – 5.2 Standard delivery)
•
Prior to any work involving the knife or the microtome, or in the cryochamber, swing the knee lever
upwards, see
.
3.3 Safety features
3.3.1 Emergency stop switch
The Leica CM3600 XP is equipped with an
Emergency stop
switch
located on the cabinet to
the left of the window.
To activate the
Emergency stop
switch:
Press the
Emergency stop
to stop the microtome immediately.
Warning
Always press the
Emergency stop
switch before making modifications to the microtome, knife or
specimen (see warning message below the window
) to prevent accidental activation of
the microtome using the knee lever.