Cleaning and Maintenance
7
7.2
Changing the fuse
Warning
Before changing a fuse, always switch off the instrument first and remove the instrument cable
completely. The instrument must have cooled down and the paraffin tank must be empty.
When changing a fuse, do NOT use any fuses other than the spare fuses supplied with the
instrument.
If the instrument fails completely, first check the power supply at the power socket.
Then check the fuses at the rear side of the instrument.
To do so, proceed as follows:
• Using a screwdriver
, carefully push out the fuse insert
• Remove the fuse insert - it contains two fuses
.
• Check that the thin wire
in the glass capillary of a fuse is intact. If not, replace the fuse
(the standard scope of delivery includes two replacement fuses).
Warning
Before plugging the power cable back in and switching on the instrument, you must have identified
and corrected the cause of the burned-out fuse.
• Insert the fuse insert with the two fuses and start up the instrument again.