BATTERY DRILL SYSTEM

User Manual 

Easy to use  
Ergonomic design  
Simple, compact battery packs with high capacity  

 

Small, robust drill with an extensive range of attachments  
Steam-sterilisable  

Date:

 

10.10.2011

 

Summary of Contents for 22/4400-00

Page 1: ...3 BATTERY DRILL SYSTEM User Manual Easy to use Ergonomic design Simple compact battery packs with high capacity Small robust drill with an extensive range of attachments Steam sterilisable Date 10 10 2011 ...

Page 2: ...rse rotation hold down both triggers R and L Specifically reverse rotation is required for procedures such as releasing bone screws and removing Kirschner wires Reverse rotation must be used exclusively for the above mentioned procedures Note The Rechargeable Battery 22 4416 00 may not be sterilized but must be treated with the same care as a surgical instrument even though it is not categorized a...

Page 3: ...ms Today health professionals and patients are at constantly increasing risk so appropriate cleaning and disinfection of these drills and attachments is extremely important If in doubt contact LEIBINGER GmbH Co KG and the manufacturer of the disinfectant and cleaning agent used Note Please observe the Instructions for use of attachments with the exception of the Rechargeable Battery 22 4415 00 22 ...

Page 4: ...to manufacturer for re No sound when trigger is pressed R or L trigger pressed Remark The lower trigger has a double function main switch and re verse rotation upper trigger clockwise rotation pair Drill cannot be switched from Solenoid switch trigger defec Reversing function does not Send to manufacturer for re reverse to clockwise rotation tive Control system defective react Continuous running o...

Page 5: ...f risk analysis Evaluation of the check list of Fundamental Requirements for this medical device Attachments Description Battery Drill System with end cap Mono Block Battery oscillating Saw Long rechargeable battery with extension Standard rechargeable battery Funnel for inserting rechargeable battery into Battery Drill System 4 807 Desktop Charger mono 240 115 V charging time 60 minutes Desktop C...

Page 6: ...nce with the provisions of MED UMW LEIBINGER GmbH Co KG will deal with disposal Key to the symbols Symbol 1 Observe instructions in User s Manual Symbol 2 Classification Type BF Symbol 3 Recyclable Symbol 4 CE 1253 in accordance with Directive 93 42 EEC Symbol 5 Protection class II device Symbol description R clockwise rotation L reverse rotation 6 ...

Page 7: ...use freshly charged batteries Preparation Step 1 Open the end cap 4 by turning in the direction indicated by the arrow Step 2 Insert the rechargeable battery 3 into the battery compartment 2 in the direction shown by the arrow at an angle of not more than 10 to 15 insert gently to avoid damaging the contacts Step 3 Replace the end cap 4 and close it by turning in the direction indicated by the arr...

Page 8: ...y the arrow Slide the chuck 2 onto the shaft 3 as far as it will go Release the chuck collar 1 Turn the chuck slightly until it engages in the shaft 3 Check the connection Hold the chuck 4 with your left hand and pull it away from the drill Fault Clutch does not engage in shaft Rectification If due to dirt clean the clutch if due to damage send to Customer Service 1 3 4 2 8 ...

Page 9: ... the Battery Drill System If it is not used there may be a loss of sterility Desktop Charger mono 240V and 115V charging time 60 minutes Please state required voltage when ordering Desktop Charger dual 240V and 115V charging time 60 minutes Please state required voltage when ordering Three jaw chuck 0 5 mm to 7 0 mm capacity for round and triangular shanks incl chuck key Reduction gears with AO st...

Page 10: ...nd attachments a short time before the operation 4 Remove blood and other exudates from instruments and attachments during the operation replace instruments and attach ments in their assigned place after use 5 Flush through cannulated instruments to prevent blood and bone powder from drying on 6 Use Ringer s lactate solution or saline solution to clean instruments and accessories but do not immers...

Page 11: ...packaging then drying is not necessary 5 LUBRICATION The joints and all moving parts of instruments and attachments must be cleaned and lubricated after use Only use biological oils and acid free lubricants pure cooking oil is recommended If special instrument oil solutions are used the instruments and attachments can be immersed for a certain period of time With some instrument washing machines t...

Page 12: ...harged the lamp glows green 2 Chargers 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 switch off automatically when the BATTERY 22 4415 00 22 4416 00 is fully charged This means that you can leave the BATTERY 22 4415 00 22 4416 00 in the operational Charger 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 for a prolonged period so that it is always charged to full capacity without any risk of overch...

Page 13: ...obtain a list detailing the type and scope of work performed with the company stamp date and signature of the repairer WARNING All equipment must be cleaned before sending for servicing Transport and storage conditions To prevent damage The Charger 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 and BATTERY 22 4415 00 22 4416 00 must be transported in the aluminium case provided which has a special...

Page 14: ... EN 529 Chargers 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 240 V and 115 V 50 HZ 14 VA 12 V 1100 mA max II IPX0 Function The Chargers 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 are controlled and monitored by a microcontroller Irrespective of the charge level of the battery the controller measures the momentary capacity and controls the charging process 14 ...