Instruction Manual
ZD200-GSM-REVA
19
6.
Disconnect the ProBus Extender Cable from the
oscilloscope. Verify that the scale factor changes from
20 mV/div to 2 mV/div.
7.
Reconnect the ProBus extender Cable to the
oscilloscope.
The warranted characteristics of the ZD200 are valid at any
temperature within the Environmental Characteristics listed in
the Specifications. However, some of the other test equipment
used to verify the performance may have environmental
limitations required to meet the accuracy needed for the
procedure. Be sure that the ambient conditions meet the
requirements of all the test instruments used in this procedure.
Note
: The correct operation of the ZD200 controls requires
software version 6.4.1.5 or higher. The software version in the
test oscilloscope can be verified by selecting
Utilities, Utilities
Setup...
from the menu bar, then the
Status
tab.
Contact your local LeCroy representative or visit
www.lecroy.com if the software in your oscilloscope requires
updating.