PP010-OM-E Rev A

 

Probe Compensation 

Proper compensation of the probe is required to assure good amplitude 
accuracy in the dynamic portions of the waveform being measured. LF 
compensation matches the probe to differences in oscilloscope input 
capacitance. The LF compensation should always be checked and adjusted as 
needed when first connecting a passive probe to the oscilloscope input. 
LF compensation is performed by connecting the input of the probe to a low 
frequency square wave, such as the oscilloscope calibrator signal set to 1 kHz. 
The compensation is adjusted by rotating the trimmer pot accessible through 
the small hole in the center of the housing near the BNC connector. Use the 
tool supplied with the probe for this adjustment. 

 

Adjusting Probe Compensation 

§ § § 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for PP010-1

Page 1: ...LeCroy PP010 Passive Probe Instruction Manual Revision A February 2006 www com information itm com 1 800 561 8187 ...

Page 2: ...rn all products under warranty with transportation charges prepaid This warranty replaces all other warranties expressed or implied including but not limited to any implied warranty of merchantability fitness or adequacy for any particular purposes or use LeCroy shall not be liable for any special incidental or consequential damages whether in contract or otherwise 2006 by LeCroy Corp All rights r...

Page 3: ...g measurement may expose you to electric Connect the probes grounding terminal to the ground potential of the device under test Failure to connect to ground potential may expose you to electric shock Connect to properly grounded instruments This probe is to be used only with test instruments having input connectors with the BNC shield ring connected to earth ground Connect the probe properly Conne...

Page 4: ...Avoid physical injury The probe tip is extremely sharp Use care when handling to prevent injury including accidental skin puncture Use of the probe and or the test instrument it is connected to in a manner not specified by the manufacturer may impair the protection mechanisms Components www com information itm com 1 800 561 8187 ...

Page 5: ... Accessory Replacement Kit PKIT2 5MM 101 contains the following items Part Quantity Standard Ground Lead 1 Sprung Hook 1 Color Coding Rings set 1 Adjustment Tool 1 Ground Attachment 1 IC Test Tip 1 www com information itm com 1 800 561 8187 ...

Page 6: ... Electrical Ratings Maximum Input Voltage 600 V DC ACpeak Note The maximum input voltage derates when frequency response is 1 MHz or higher See voltage derating curve on the next page EC Declaration of Conformity The PP010 probe meets intent of the European Council Directive 73 23 EEC for product safety per the following standard EN 61010 031 2002 Safety requirements for hand held probe assemblies...

Page 7: ... Derating Curve 10 100 1000 0 1 1 10 100 1000 Frequency MHz Voltage V DC AC peak Typical Input Impedance 10 100 1000 10000 100000 1000000 10000000 0 01 0 1 1 10 100 1000 Frequency MHz Zin ohms www com information itm com 1 800 561 8187 ...

Page 8: ...djusted as needed when first connecting a passive probe to the oscilloscope input LF compensation is performed by connecting the input of the probe to a low frequency square wave such as the oscilloscope calibrator signal set to 1 kHz The compensation is adjusted by rotating the trimmer pot accessible through the small hole in the center of the housing near the BNC connector Use the tool supplied ...

Reviews: