14 Re-issuing Leckey
Products
Most Leckey products are assessed and
ordered to meet the needs of an individual
user. Before reissuing a product we
recommend that the therapist prescribing
the product has carried out an equipment
compatibility check for the new user and has
ensured that the product being re-issued
contains no modifications or special
attachments.
A detailed technical inspection should be
carried out on the product prior to re-issuing.
This should be carried out by a technically
competent person who has been trained in
the use and inspection of the product. Please
refer to section 13 for the required checks to
be carried out.
Ensure the product has been cleaned
thoroughly in accordance with section 11 of
this manual.
Ensure a copy of the user manual is supplied
with the product. A copy can be downloaded
from our website www.leckey.com
Leckey recommend that a written record is
maintained of all product inspections carried
out during the reissue of the product.
If in any doubt to the continued safe use of
your Leckey product or if any parts should
fail, please cease using the product and
contact our customer service department or
your local dealer as soon as possible.