Notes
1.
The temperature measuring equipment should be used in the room between 18
°
c and 35
°
c at room
temperature. Do not install the temperature measuring equipment under the air vent to ensure that there is
no heating source within 3 meters.
2.
When a person enters the room from a cold outdoor environment, the temperature measurement accuracy
will be affected. The forehead temperature test shall be conducted after the forehead temperature is stable
and the forehead is not shielded for three minutes.
3.
The temperature read by the temperature measuring device is the temperature of the forehead temperature
area. When there are water, sweat, oil stains on the forehead, or heavy makeup or wrinkles in the elderly,
the temperature read will be lower than the actual temperature to ensure that there is no hair or clothing
cover in this area.
4.
Temperature measuring equipment shall not be placed in a humid environment. Outdoor equipment shall be
provided with a damp proof cover. Wet use environment will accelerate the rapid aging of electronic
components, thus reducing the service life of the product, resulting in moisture damage to parts.
Thermal imaging, body temperature screening compared with professional medical level
temperature measuring method, the contact-less temperature measurement, safety non-inductive,
and the advantages of fast and efficient, is for the urban traffic, schools, companies, offices,
business to wait good solutions have been supervised rapid temperature screening. In addition, the
thermal imaging temperature cervix on installation environment, detection distance has certain
requirements, such as in the use of thermal imaging when the cervix temperature, suggested in a
relatively stable indoor environment, temperature measuring range, required by original factory
for calibration. All test distance must be in the calibration position shall prevail. All products
installed in accordance with the requirements of manufacturer must be in indoor constant
temperature environment.