Follow the steps a) – g) as discussed above except that you define one waste
position rather than a group. The ‘Type’ must be ‘Reagent’.
FIGURE B-8 The Group Editor
B.4
Electrical Connections
To control the TTL inputs of the Probot (Section 2.4.1 D) by the relay outputs of
the UltiMate Micropump:
a)
Connect the event output 3 (Open Collector 1) to the Probot START input as
presented in TABLE B-1.
b)
Connect the event output 7 (relay 1) to the Probot TTL 1 (‘Peak start’) input
as presented in TABLE B-1.
c)
Connect the event output 8 (relay 2) to the Probot TTL 2 (‘Peak end’) input
TABLE B-1
Connecting the Probot and the Micropump
Probot
Color Code
UltiMate Micropump EVENTS
Start
Signal
Start
Lilac
Event 3, pin 3
Ground
Black
Ground, pin 4
TTL Inputs (TT1=Peak Start, TTL2=Peak End)
TTL1, pin 1
Red
Event 7, pin 7
TTL2, pin 3
Orange
Event 8, pin 9
Ground, pins 9,11,13,15
Blue
Event 7 and 8, pins 8, 10
B.5
Listing of the CHROMELEON Program used for this Example
Pressure.LowerLimit =
0.0
Pressure.UpperLimit =
400.0
%A.Equate =
"%A"
%B.Equate =
"%B"
%C.Equate =
"%C"
%D.Equate =
"%D"
CRP =
104.2
ParkPercentage =
Disabled
Summary of Contents for Probot 161403
Page 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Page 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Page 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Page 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Page 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...