Appendix B
User’s Manual Probot
D960R3
B-3
B.3
Setting up µ-Carrier for Peak Collection
The following sections describe how to setup the
µ
Carrier (V2.0) software for
peak collection. The Probot Micro Fraction Collector is controlled via external
contact signals (START, TTL inputs). The detected peaks are collected in
collection vessels and the mobile phase in between peaks can either be collected
into one waste vessel or into a group of waste vessels.
To setup
µ
Carrier for peak collection using a waste group:
a)
Prepare a collection group with the amount of vessels required to collect all
peaks and a waste group with the appropriate amount and size of vessels
(FIGURE B-2).
FIGURE B-2 Defining the Collection, Waste and Park Groups
Note: The number of vessels required to collect the peaks and the waste may
differ and depends on the chromatography conditions (e.g. the peak resolution)
and the trigger settings.
b)
If required, define a park position and a wash position. FIGURE B-3 shows all
groups used for the peak collection task in this example.
FIGURE B-3 Collection, Waste and Park Groups
Waste
Group
Collect
Group
Park
Position
Summary of Contents for Probot 161403
Page 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Page 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Page 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Page 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Page 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...