Appendix D
User’s Manual Probot
D960R3
D-5
D.4
Setting up µCarrier and CHROMELEON
®
The following sections provide basic information of how to setup µCarrier
software to control the waste removal option properly. In addition, an example is
provided which shows usage of this option in conjunction with the peak
detection capabilities of CHROMELEON. For more detailed information refer to
the manuals provided with the software packages and APPENDIX B.
D.4.1
Enabling the Vacuum Waste Removal Option
The dosage module cannot be used in conjunction with the waste removal
option. Select the
Controls
tab in the
Settings
menu; then disable the ‘Syringe
pump module …’ (dosage module). Enable the ‘Vacuum waste removal’ option
(FIGURE D-5).
FIGURE D-5 Disable the Dosage Module
D.4.2
Adjusting the Fused Silica Capillary and the Needle Height
The following steps are required to adjust the fused silica capillary properly in the
hollow stainless steel needle.
a)
Select the
Contact
tab and select the
NeedleUp
button to make sure that the
needle is in its upper position (FIGURE D-6).
FIGURE D-6 XYZ Robot Section – Contact Tab
b)
Check that the fused silica capillary extends the needle tip by approximately
4 mm (0.0016 inch, FIGURE D-2).
Summary of Contents for Probot 161403
Page 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Page 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Page 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Page 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Page 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...