10
start countdown.
6.
The countdown settings are independent of the temperature settings. When
setting the countdown, take into account the time required to achieve the set
temperature.
The countdown timer setting as following procedures: For example:
Set the countdown time from 0 to 10 min.
Operation
Display
°C
Select the menu option
Cdt
Navigate to the countdown timer selection menu
option
OFF
Press set key to enter the countdown timer setting
If the countdown time is 0, it displays OFF.
If not, it displays the countdown timer value
010.
Set the countdown time you require
The “.” In the lower right corner
means that it is in
countdown timer setting
9.57
Press set key to save the setting
The countdown timer will begin once you save the
setting
The setting will exit automatically after 5S without
operation
5.8
Setting the dry-start protection alarm
When the dry-start protection function is on, the bath will stop heating and an
alarm will sound when the system detects that the inner chamber is in dry-start
status. When the dry-start protection is off, there is no protection alarm and the
bath will continue heating to the set temperature.
For example:
the current temperature is 37.0°C, set the dry-start protection from
“ON” to “OFF”;
Operation
Display
°C
Select the menu option
OPA
Select the dry-start protection setting menu option
On
Press set key to enter the dry-start protection
setting. The default setting is “ON”
Summary of Contents for WB-Pro