background image

 

17 

flasks, a 2 ml pipette, a 10 ml pipette, six 25 ml test tubes, six round or square 
cuvettes, and a hot plate. 
 
Procedure: 
1. 

Fill the burette to the top calibration line (50 ml) with the glucose stock 

solution.  Deliver 5 ml of the glucose solution from the burette into the 
volumetric flask.  Dilute the flask to the mark with de-ionized water, cape the 
flask, and mix the diluted solution thoroughly.  Transfer this first standard to 
an Erlenmeyer flask, calculate its concentration, and label the flask.  Rinse 
the volumetric flask with de-ionized water and repeat step 2 for standards 2, 
3, and 4 using 10 ml, 15 ml, and 20 ml of the stock solution. 

4. 

Pour 10 ml of the soft drink into the 100 ml volumetric flask, swirl it until 
bubbling stops, and dilute to the mark.  Transfer this solution to an 
Erlenmeyer flask.  Rinse the volumetric flask, pipette 10 ml of the diluted soft 
drink solution into it, dilute and mix.  Transfer this solution to another 
Erlenmeyer flask.  Rinse the pipette and volumetric flask.  Pipette 10 ml of the 
second dilution into the volumetric flask, dilute and mix.  This final unknown 
solution is the 1:1000 dilution of the original soft drink. 

5. 

Label six test tubes.  Fill them with 25 ml of water using a pipette, mark the 
solution level, and empty them.  Pipette 2 ml of water, the unknown and the 
four standards into the tubes, respectively.  Rinse the pipette after each use. 

6. 

Using a graduated cylinder, add 2 ml of the alkaline copper titrate solution to 
each test tube and stir by swirling the tubes.  Rinse the graduated cylinder 
well. 

7. 

Fill half of a 600 ml beaker with water and place it on a hot plate to boil.  
Prepare an ice-water bath in a second 600 ml beaker. 

8. 

Immerse the six tubes into the boiling water for exactly six minutes to oxidize 
the glucose.  Remove the tubes from the boiling water and immerse them in 
the ice water bath for exactly 3 minutes to allow them to cool. 

9. 

Using a graduated cylinder, add 2 ml to the phosphomolybdic acid solution to 
each tube and stir by swirling them.  Allow the solution two minutes to react, 
and then dilute them to the 25 ml mark with de-ionized water. 

10. 

Place rubber stopper in each tube and mix by inverting the tubes several 
times. 

11. 

Set the wavelength to 780 nm and put the second order filter lever in the red 
position.  Fill a cuvette with a blank solution and insert it into the sample 
compartment.  Blank the instrument according to the procedure in 

Basic 

Operation

12. 

Fill another cuvette with the first standard and measure its absorbance on the 
spectrophotometer. 

 
Repeat 

Step 12

 for the other three standards and the unknown solution. 

Calculation: 
1.  On a piece of regular graph paper, label the horizontal axis concentration and 

mark it in equal intervals from 0 to the value of standard 4.  Label the vertical 

Summary of Contents for Alpha-1000

Page 1: ...Alpha 1000 SPECTROPHOTOMETER USER S MANUAL V 1 0 Revision 0 Laxco Inc http www laxcoinc com...

Page 2: ...2...

Page 3: ...s at Different Wavelengths 10 Lamp Changing 11 Lamp Adjustment 12 Absorbance Accuracy and Stray Light Checking 13 Sample Experiments 13 Experiment 1 A Brief Introduction to Beer s Law 13 Experiment 2...

Page 4: ...lay to show absorbency and transmittance Figure 1 illustrates the relationship between these parts Light Source Monochromator Sample Compartment Detector Digital Display Figure 1 Block Diagram for the...

Page 5: ...adiant Energy 2 T at 400 nm Photometric Range 0 T to 125 T 0 3A to 2 0A Photometric Accuracy Better than 2 0 T Data Output USB RS232 Power Requirements 85 240 VAC 50 60 Hz Dimensions 455 L x 340 W x 1...

Page 6: ...uency fields Set the unit up in an area that is free of dust corrosive gases and strong vibrations 2 Remove any obstructions or materials that could hinder the flow of air under and around the instrum...

Page 7: ...It is used for connecting to Computer with Application Software RS232 It is used for Firmware upgrading Figure 2 Alpha 1000 Spectrophotometer Sample Compartment Digital Display 0A 100 T Knob Operatio...

Page 8: ...Parts List of this manual for details Follow the steps below to change the Cuvette Sample Holder Open the lid of the Sample Holder and locate the Sample Holder Locking Screw as shown in Figure 4 Use A...

Page 9: ...solution will typically be distilled water but can be the aqueous in water medium that you are using for your experiment The transmittance of this solution is made to read 100 T 0A This process is ca...

Page 10: ...ent this mark is located at the front of the sample compartment Be sure that the test tube has been firmly pressed into the sample compartment and the lid of the sample compartment has been closed 4 A...

Page 11: ...the lamp housing in place refer to the diagram shown below 3 Remove the screw and retain BE CAREFUL THE BULB MAY BE HOT 4 Remove the old bulb and place the leads of the new bulb into the holes in the...

Page 12: ...light bulb is fully inserted into its holder The bulb socket has been permanently mounted in the optimal position by the factory So the rest of steps should not be needed 3 The lamp socket is mounted...

Page 13: ...m s the speed of light in a vacuum often rounded to 3 00 x 10 8 m s and wavelength of light Every substance has a characteristic absorbance of the light spectrum which means that there are wavelength...

Page 14: ...ally zero It is found that absorbance is directly proportional to concentration so that A EBC Where A The absorbance C The concentration in moles liter and B The width of the sample cell or cuvette in...

Page 15: ...sure that the meter display 0 T and 100 T readings are adjusted at each new wavelength Calculations 1 Study the data and identify the points where the peaks are likely to be found Make measurements a...

Page 16: ...f the first standard at this wavelength 6 Repeat step 5 for the other three standards Calculations 1 On a sheet of graph paper label the horizontal axis concentration and mark it in equal intervals fr...

Page 17: ...known and the four standards into the tubes respectively Rinse the pipette after each use 6 Using a graduated cylinder add 2 ml of the alkaline copper titrate solution to each test tube and stir by sw...

Page 18: ...ne hydrochloride solution 10 ml 1M NaC2H3O2 aq sodium acetate solution 10 ml 1 w v 1 10 phenanthroline ortho solution a 100 ml beaker a 50 ml burette five Erlenmeyer flasks a 50 ml graduated cylinder...

Page 19: ...pared unknown solution may not yield an absorbance value that can be properly interpreted from the standard curve Therefore you may have to make adjustments in the concentration of iron in the unknown...

Page 20: ...5 COD Vial Sample Holder A 110 220 Power Cord European plug A 150 115 Holmium Oxide Filter 10 mm x 10 mm x 45 mm requires A 110 103 A 150 116 Didymium Filter 10 mm x 10 mm x 45 mm requires A 110 103 G...

Page 21: ...cannot be set to 00 0 T Sample holder Remove Cuvette Adapter or test tube Sample holder shutter May be stuck open Close shutter Defective electronic component Call an authorized service engineer Inco...

Page 22: ...Defective or dirty detector or defective electronic component Call an authorized service engineer Incorrect readings obtained Insufficient sample volume Fill cuvette with more sample Wrong wavelength...

Reviews: