- 20 -
INPUTS
Default: input 1 =
input 2 =
input 3 =
Possible operation modes:
-
(NET/GROSS): by closing this input for no more than one second, it’s making an
operation of SEMI-AUTOMATIC TARE and the display will show the net weight. To display the
gross weight again, hold the NET/GROSS input closed for 3 seconds.
-
: by closing the input for no more than one second, the weight is set to zero (see section
WEIGHT ZERO-SETTING FOR SMALL VARIATIONS (SEMI-AUTOMATIC ZERO)
).
-
: keeping the input closed the maximum weight value reached remains on display. Opening
the input the current weight is displayed.
-
: closing the input no operation is performed, the input status may however be read remotely
by way of the communication protocol.
-
: closing the input for max one second the weight is transmitted over the serial
connection according to the fast continuous transmission protocol only once (
only if
is
set in the item
)
.
-
: when the input is closed the weight is displayed based on the set coefficient (see setting
of the units of measure and coefficient), otherwise the weight is displayed.
-
:
when the input is closed the data are sent for printing if in the communication protocol
of either serial port the parameter
is set.
-
: when the input is open, the alarm
is displayed; the weight cannot be saved
(see section
LIMIT MODE
).
LIMIT MODE
The menu is displayed only when one of the inputs is set as LIMIT
.
(from 1 to 65535; default: 10): the time value, in milliseconds, that the instrument lets pass
before recognizing the closure of the input.
Example: by setting 100, the instrument recognizes the input as closed after 0.1 seconds from the
actual closure; if during this time laps the input should open and close again, the time counting
would start all over.
(from 1 to 65535; default: 1000): the time value, in milliseconds, that the instrument lets pass
before recognizing the opening of the input.
Example: by setting 3000, the instrument recognizes the input as open after 3 seconds from the
actual opening; if during this time laps the input should close and open again, the time counting
would start all over.
+