V1
Versafreeze Chest Freezers
21 / 42
8
Alarm limit values
Setting a limit value for the cover alarm delay
Tap the Menu symbol on the touch screen unit. In the parameter level, scroll down to the USr user level, tap parameter A6,
set the required limit value, and save. A timer starts when the cover is opened.
If the timer exceeds the preset delay time, the door alarm is triggered.
Overtemperature limit value
Tap the Menu symbol on the touch screen unit. In the parameter level, scroll down to the USr user level, tap parameter A13,
set the required limit value, and save. If the refrigeration compartment temperature exceeds the upper preset limit value,
the overtemperature alarm triggers.
Low temperature limit value
Tap the Menu symbol on the touch screen unit. In the parameter level, scroll down to the USr user level, tap parameter A15,
set the required limit value, and save. If the refrigeration compartment temperature falls below the lower preset limit value,
the low temperature alarm triggers.
USr User level
A 6
= Alarm delay (door open)
Factory setting
60 sec.
A 13 = Lower limit 1 (abs/rel)
Factory setting
-4.0 K
A 15 = Upper limit 1 (abs/rel)
Factory setting
4.0 K
C 11 = Refrigeration compartment temperature set point
The value is stored on the device card
C 25 = Hysteresis sensor F1
Factory setting
The value is stored on the device card
H 11 = Offset correction sensor F1
Factory setting
The value is stored on the device card
The device card is located behind the right side cover of the machine compartment of the VF 200.. C and below the up-
per cover of the machine compartment of the VF 550.. C and VF 750.. C.
Summary of Contents for VF 20040 C
Page 2: ...2 42 Versafreeze Chest Freezers V1...
Page 4: ...4 42 Versafreeze Chest Freezers V1...
Page 27: ...V1 Versafreeze Chest Freezers 27 42 VF 20040 C 115V_60Hz...
Page 28: ...VF 20040 C 230V_50Hz...
Page 29: ...V1 Versafreeze Chest Freezers 29 42 VF 20085 C 115V_60Hz...
Page 30: ...VF 20085 C 230V_50Hz...
Page 31: ...V1 Versafreeze Chest Freezers 31 42 VF 55040 C_VF75040 C 115V_60Hz...
Page 32: ...VF 55040 C_VF75040 C 230V_50Hz...
Page 33: ...V1 Versafreeze Chest Freezers 33 42 VF 55085 C_VF75085 C 115V_60Hz...
Page 34: ...VF 55085 C_VF75085 C 230V_50Hz...
Page 35: ...V1 Versafreeze Chest Freezers 35 42 Touch screen unit Single board cryostat A1...
Page 38: ...14 Notes...
Page 40: ...16 Product Returns and Clearance Declaration...
Page 41: ...V1 Versafreeze Chest Freezers 41 42 17 EC Declaration of Conformity...