4
Before starting up
4.1 Placement
Very specific placement conditions are applicable to the devices.
These placement conditions are mainly specified in the technical
data for the device.
Further information about the technical data can be
found at
Chapter 11.1 ‘General data’ on page 93
Additional placement conditions are described below.
n
Irritant vapours can be produced, depending on the heat
transfer liquid used and the operating mode. Ensure adequate
extraction of these vapours.
n
Observe the requirements of the device for electromagnetic
compatibility (EMC).
n
Do not cover the ventilation openings.
Further information about EMC requirements can be
found at
Chapter 1.5 ‘EMC requirements’ on page 8
Operation outdoors at outdoor temperatures below
5 °C
A warning is displayed on the screen indicating the
duration of the preheating time for the compressor and
whether the compressor needs preheating. Increased
wear or damage to the compressor may result if the
compressor is not preheated! Further information can
be found at
‘Outdoor installation’ on page 21
.
Personnel:
n
Operating personnel
WARNING!
Rolling or toppling of the device from incorrect han-
dling
Impact, crushing
Do not tilt the machine.
Place the device on a level, skid-free surface with
sufficient load bearing capacity.
Engage the reel brake when setting up the device.
Do not place any heavy parts on the machine.
1.
Place the devices on a suitable base.
The devices can be moved. Release the locking
brakes of the castors for this by pushing the lever
upwards.
Before starting up
Variocool
23
Summary of Contents for VC 1200