84 / 137
Proline Kryomats
7.11.3
Internal and external control parameter sets
If a thermostat is used for a number of applications, which always leads to a change of the control parameters, these
control parameters (up to 9 sets) can be saved in the thermostat and activated again as required.
Also saving is useful for finding the best control parameters; in this way external management of the control
parameters can be avoided.
There are 9 sets (each for internal and external sets of control parameters) saved at the factory.
In this menu the control parameters cannot be edited, they are only displayed.
With
Activate
the currently valid control parameters are used.
With
Upload actual
the actual ones are read in and saved (for later reuse).
With
Default
the set of control parameters saved at the works is loaded again (in this case the control parameters
set by the customer are lost).
Command
Control parameter sets
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to:
Control
Control
Parameters
Control para. sets
.
The adjacent window appears.
Set 1 to Set 9.
Select the desired set with
and confirm
with
.
Select the desired set to be changed with
and confirm with
.
In the setting window (see left) the selected set
is listed under
internal
or
external
in the
display.
Under
Status
the previously selected set:
is activated,
is read in and
the set, which was saved at the factory, is
restored.
Pump Menu
End
T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Correction limitation
Set 1
Set 2
Set 3
Set 4
Set 5
Set 6
Set 7
Set 8
Set 9
Pump Menu
End
T
set
T
fix
Status
intern
extern
Activate
Upload actual
Default