62 / 137
Proline Kryomats
7.6.5
Setting the volume of the acoustic signals
The LAUDA Proline Thermostats signal alarms as a dual-tone acoustic signal and warnings as a continuous tone.
Command
Sounds
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to
Settings
Basic Settings
Sounds
.
Select either
Alarm
or
Warning
.
Example on left:
Alarm
is set to loud.
Select the desired volume with
or
.
Accept selection with
or End or quit the
window with without making changes.
7.6.6
Entering the offset of the internal temperature probe
If, during checking with a calibrated reference thermometer a deviation is found, then the offset (i.e. the additive part
of the characteristic) of the internal measuring chain can be adjusted with the following function. The reference
thermometer must be dipped into the bath according to the details on the calibration certificate.
Command
Calibration
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to
Settings
Calibration
intern Pt100
Calibration
.
The window shown on the left appears.
Confirm selection with
.
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Pump Menu
End
T
set
T
fix
loud
medium
low
off
Alarm