48 / 137
Proline Kryomats
7.5
Important settings
7.5.1
Temperature setpoint setting
The setpoint is the temperature, which the thermostat should reach and maintain constant.
Master (main level)
SEt
Press key until
SEt
(Setpoint) appears.
Press key, display flashes.
or
Enter the setpoint with the two keys (
7.4.1 General key functions and
Wait 4 seconds or
Display flashes 4 seconds
new value is automatically accepted, or
value is accepted immediately with Enter key.
For safety reasons the setpoint can only be set up to 2°C above upper limit
of the operating temperature range for the relevant device type.
In the following cases, the manual setpoint entry is blocked:
Setpoint is taken from the analog module, from the programmer in the
Command remote control or via the serial interface.
When the setpoint temperature is to be lowered, it may take up to one
minute before the blue LED lights.