Proline Low-temperature Thermostats
40
Starting up
YACE0072 / 21.08.07
7.6.5 Submenu
LMDkt
Æ
O0c--
(Master): Setting internal control parameters
You only obtain the following possible settings when the control variable is set to "Internal" (
⇒
Section
Set control parameters
O0c--
°C
Display flashes. Set
with or .
Xp range: 0.3K … 25.0 K.
Proportional range [K]
^O &/
°C
* Important:
The windows for setting the
internal control:
rs
(=tv) and
rc
(= td)
are only shown when
rsc
(=tvd) has been changed
from the works setting
"Automatic" to "Manual".
Changeover should only be
carried out by experienced
control technicians.
Display flashes. Set
with or .
Tn range: 3s … 180s, Off.
Reset time [s]
rl 2/
°C
*Display flashes. Set
with or .
Tv range: Off, 1s … 145s.
Deriv. time [s]
rs 10
°C
*Display flashes. Set
with or .
Td range: 0.0s … 25s.
Damping time [s]
rc §4
°C
longer
than
3 sec.
Display flashes. Change tv
and td from
@
=Autom. to
LM
=Manual
with or .
Auto/manual for tv, td
rsc @
°C
The works setting (default)
of the internal control
parameters has been
restored and
cmlD
is
displayed.
longer
than
3 sec.
Control works settings
cDE
°C
End
Dlc
°C