Alpha
29 / 52
7.8
Offset for the internal temperature probe
CAL
Keep attention: All change done here cause the loss of factory calibration!
Calibration should only be activated if a sufficiently accurate temperature measurement device is available as reference.
If, during checking with a calibrated reference thermometer, (for example from the LAUDA DigiCal Series), a devia-
tion is found, then the offset (i.e. the additive part of the characteristic) of the internal measuring chain can be adjusted
with the following function. The reference thermometer must be dipped into the bath according to the details on the
calibration certificate. Only offset values of
±
5 Kelvin are accepted. Switching on the unit, the last used CAL value is
preloaded again.
5x
to
CAL
and 1x
longer than
3 seconds.
Call the option
CAL
.
or
Display flashes. Enter the value of the reference thermometer.
The selected value is displayed flashing.
Wait 4 seconds or
Display flashes 4 s
new value is accepted automatically,
or
value is immediately accepted with the <Enter> key.
7.9
Restoring works settings
dEF
If you would like to restore all the works settings except the probe calibrations
CAL
…
6x
to
dEF
and 1x
call the works settings
dEF
.
dEF
is flashing.
3 seconds long
Press <Enter> more than 3 seconds (display flickering).
After button release the changeover is acknowledged with
donE
, the unit
returns back to the mode “Bath temperatur”.
Reset Values: Set = 25 °C, Hi = 105 °C, Lo = -30 °C, Cool = Auto
d E F
Summary of Contents for RA 12
Page 12: ...12 52 Alpha 7 2022 1 Control head with clamp 2 Bath vessel 3 Four mounting feet 1 2 3...
Page 14: ...14 52 Alpha 7 2022 1 Mains cable 2 Rating label 3 Cooler cable 4 Bath drain nozzle 1 2 3 4...
Page 44: ...44 52 Alpha 7 2022 Cooling characteristics Alpha RA 8 RA 12 and RA 24 230 V 230 V...
Page 45: ...7 2022 Alpha 45 52 11 Declaration of Conformity and Product Returns...
Page 46: ...46 52 Alpha 7 2022...
Page 47: ...7 2022 Alpha 47 52...
Page 51: ......