Proline Edition X heating thermostats
YACE0099 / 10/9/2018
Interface modules
93
START
Switches the device on (from Standby).
See safety information
(
STOP
Switches the device into Standby (pump, heater off)
RMP_SELECT_X
Selection of the programme (1
— 5) to which the further instruc-
tions apply. When the unit is switched on, programme 5 is se-
lected automatically.
RMP_START
Start the programmer
RMP_PAUSE
Hold (pause) the programmer
RMP_CONT
Restart the programmer after pause
RMP_STOP
Terminate the programmer
RMP_RESET
Delete the programmer (all Segments)
RMP_OUT_00_XXX.XX_XXXXX_
XXX.XX_X
Set a programme segment (temperature and time, tolerance and
pump level). A segment is added and appropriate values are
applied to it.
RMP_OUT_02_XXX
Number of times the programme runs: 0 = unlimited / 1
— 250.
Fo
r “_“ use also “ ” (blank character).
Response from thermostat “OK“ or in case of error ”ERR_X“ (RS-485 interface
e.g. “A015_OK” or in case of error ”A015_ERR_X”.).
Permitted data formats:
-XXX.XX
-XXX.X
-XXX.
-XXX
XXX.XX
XXX.X
XXX.
XXX
-XX.XX
-XX.X
-XX.
-XX
XX.XX
XX.X
XX.
XX
-X.XX
-X.X
-X.
-X
X.XX
X.X
X.
X
-.XX
-.X
.XX
.X
8.3.6 Read commands (Data requested from the thermostat)
Command
Explanation
IN_PV_00
Read bath temperature (outflow temperature)
IN_PV_01
Indication of the controlled temperature (internal/external Pt/external
Analogue/external Serial)
IN_PV_03
Read external temperature TE (Pt100)
IN_PV_04
Read external temperature TE (Analogue input)
IN_PV_05
Read current bath level
IN_PV_10
Read bath temperature (outflow temperature)
in 0.001 °C
IN_PV_13
Read external temperature TE (Pt100)
in 0.001 °C
IN_SP_00
Read temperature setpoint
IN_SP_01
Read current pump power level
IN_SP_02
Read cooling operation mode (0 = OFF / 1 = ON / 2 = AUTOMATIC)
IN_SP_03
Read current overtemperature switch-off point
IN_SP_04
Read current outflow temperature limit TiH
IN_SP_05
Read current outflow temperature limit TiL
IN_PAR_00
Read control parameter Xp
IN_PAR_01
Read control parameter Tn (181 = OFF)
IN_PAR_02
Read control parameterTv
IN_PAR_03
Read control parameter Td
IN_PAR_04
Read control parameter KpE
IN_PAR_05
Read control parameter TnE (980 = OFF)