Proline Edition X heating thermostats
YACE0099 / 10/9/2018
49
Enter the desired limit temperature.
Accept the change with
,
or quit the window with
without making
changes.
7.8.4 Setpoint offset operating mode
With this function it is possible to apply an offset value to the temperature provided by the external tem-
perature probe or a module and then to use it as the setpoint. The bath temperature can, for example,
be operated at -25 °C below the temperature of a reactor, which is being measured by the external tem-
perature probe.
Master
SEtq.
3x
to
MNEnu
to
PArA
5x
SEtq.
in the following submenu the relative setpoint (Set) and
the source of the actual value can be entered.
Continue with
.
The offset operating mode is currently deactivated.
To activate it press
and with change to
EP
(external Pt100),
EA
(external via analog module) or
ES
(external via serial mod-
ule).
Wait 4 seconds or
Display flashes 4 s
new value is automatically accepted,
or
value is immediately accepted with the Enter key.
Continue with . The left-hand window is displayed.
To activate the relative offset input press
and change the value
with or
.
Wait 4 seconds or
Display flashes 4 s
new value is automatically accepted,
or
value is immediately accepted with the Enter key.
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Lower limit (Til)
-50,
0
Min: -50,0 °C Max: 301,0 °C
SEtq.
°C
r 0FF
°C
r0FFS
°C