Contact module
8.5.1 Contact module LRZ 915 with three inputs and three outputs
Contact module (order no. LRZ 915) on 15 pole SUB-D socket. With three relay contact outputs
(changeover, max. 30 V/0.2 A) and three binary inputs for control via external voltage-free contacts.
The following functions are made available by the inputs:
-
Set fault with function: Master:
F ALA
or Command:
Fault
.
-
Set Standby with function:
F Stb
or
Standby
. See safety information (
þ
7.9.3).
-
Control programmer (Input 1 activates programmer 1, input 2 activates programmer 2 etc.
At the first “close” the programmer gets starting, “open” removes it in “pause”. The next
“close” initiate “continue”) with function:
F Pr6
or
Programmer
.
-
Control alternating mode (the switching state contact “open“ or “closed” allot to two differ-
ent set point temperatures) :
F t2C
or
alternating mode
.
-
Controller mode the switching state input “open” or “closed” can allotted to two different
control temperature sources. E. g. internal
↔
external control):
F Con
or
type of control
.
The following functions are made available by the outputs:
-
Signal various fault states:
F diA
or
fault diagnosis
.
-
Signaling standby:
F Stb
or
Standby
.
-
Providing status of the window discriminators (inside
↔
outside):
F WXi
or
temperature range
.
-
Providing the programmer status:
F Pr6
or
Programmer
.
-
Signalling refill of heat transfer liquid:
F F1L
or
Refill
.
Pump Menu Screen
T
set
T
fix
0,40bar
Pumpe
6
T
set
20,
00°C
T
ext
---,
- -°C
Niveau
9
0,0
Y(%)
-5,5
T
out
°C
55,
00
T
ret
0,
00°C
Summary of Contents for Integral XT 150
Page 2: ......
Page 4: ......
Page 164: ...Integral XT 162 YAWE0028 24 06 2016 Intentionally left blank ...
Page 166: ......
Page 167: ......