10
Faults
10.1
Alarms, errors and warnings
All alarms, error messages and warnings triggered on the device appear in
text form on the display.
Alarms affect safety. The components of the device, such as the pump,
switch off. The device emits an sound. Once the cause of the fault has been
eliminated, the alarm can be canceled with the Enter key.
Refer to
Chapter 10.2 “Alarms” on page 147 for a list of alarms.
Warnings do not have a significant effect on safety. The device continues
to operate. The device will make a continuous noise for a short period of
time. Warnings are issued periodically. Once the cause of the fault has been
eliminated, the warning can be canceled with the Enter key.
If an error occurs, the device emits a sound.
If this happens, switch off the device at the mains switch. If the error occurs
again after switching on the device, make a note of the error code and the
corresponding description and contact the LAUDA Constant Temperature
Equipment Service department. You will find the contact information here
Chapter 14.4 “Contact LAUDA” on page 160.
Errors are displayed with an appropriate description and an error
code in the form of a consecutive number.
10.2
Alarms
Alarms are shown on all displays in use.
Table 41: Constant temperature equipment alarms
Code
Message
Description
User action
1
Low-level pump
Pump detects low level
Refilling the heat transfer liquid
2
Low level
Float detects low level
Refilling the heat transfer liquid
3
Overtemperature
Overtemperature (T > Tmax)
Allow the device to cool down to
T < TMax; adjust TMax if necessary
4
Pump blocked
Standstill of the pump
Switch off the device, check the vis-
cosity
5
Verb. Command
Command Touch remote control with-
drawn during operation.
Plug in the cable of the Command
Touch remote control unit.
6
---
---
----
Procedure in event of alarm
Procedure in event of warning
Procedure in event of error
V6
Integral Process Thermostats and High-Temperature Thermostats
147 / 198