12 / 34
Hydro Water Baths
09/2020
8.1.1
Setting the required temperature
In this input window, the required temperature of a process is set. The temperature can be set between 10
°C and 100 °C. The setting range of the required temperature can be limited within the values pre-set in
the basic setting menu (see 8.2.1) and is displayed in either °C or °F (see 8.2.3). Set point 100 °C is a
pure boiling mode. The Water Bath's heating element is supplied with energy constantly.
Use keys
①
and
②
to set the required temperature value, confirm with key
③
and
return to the main menu. The set point is now displayed beneath the temperature sym-
bol in the main menu.
Key
①
Key
②
Key
③
Factory setting 50.0 °C
8.1.2
Setting the switch-on delay
In the opened input window, the value t in hours and minutes (h:min) for delayed switch-on of the heat-
ing can be pre-set. The delay can be set between 0:00 (operation without delay) and max
999:59 h:min.
Set the required time value with keys
①
and
②
, then confirm with key
③
and return to
the main menu. The set delay will be displayed beneath the symbol of the switch-on
delay in the main menu.
Key
①
Key
②
Key
③
Factory setting 0:00 h:min