ECO GOLD
54
Maintenance
25/08/2011/ YACE0088
8.2
Device status
Here, accumulated error messages as well as device and software data can be recalled.
−
Access to the main menu level is obtained by pressing the
enter key
.
−
Selection and confirmation of
Æ
Setup
Æ
Device Status
.
The adjacent menu window appears.
Here, you can now
−
Errorstore
Read out the error store
−
Device data
Request device data
−
SW version
Request the software version
−
Type
Request the device type
−
Serial numbers
Request serial number.
8.2.1
Store for errors, alarms and warnings
For error analysis the ECO thermostats have an error store in which up to 140 warning, alarm and error
messages can be saved.
Error store
−
confirm
with .
The latest message is located in the first position.
−
You
navigate
with or through the results which are
sorted by date. The message text appears in the footer.
The relevant module which is causing the message is dis-
played under "Source".
"Code" is the coded alarm, warning or error description.
"Type" specifies alarm, warning or error. The list of alarms and
warnings can be found in the appendix (
Ì
8.2.2
Device data
Device data
−
confirm
with .
The device parameters are displayed under the menu point
Device data.
ESC
סּ
OK
Standby
Device status
Errorstore
►
Device data
►
SW version
►
Type RE 420
►
Serial numbers
►
No
.
Sourc
e:
Code Type Date
Time
5
Con-
trol
29 Error 30.10.09 10:32
4
Safety 3
Alarm
30.10.09
10:32
3
Con-
trol
4 Warn 29.10.09 16:41
2 Safety
29 Error 28.10.09 17:02
1
Con-
trol
36 Error 28.10
08:04
Safety
Overtemperature
ESC
סּ
OK
Standby
Device data
Tint
22.23°C
T_lp
-86.33°C
T_b_s
23.04°C
T_a
-96.33°C
T_max_s
85°C
T_mos
30.93°C
Pump
power