26 Operator’s Manual
The following procedure is to be used to top off the water level for the
GreenLight HPS
™
laser system only if a low water fault occurs.
1.
Turn the circuit breaker off and unplug the laser.
2.
Remove the reservoir cap in the rear of the laser.
3.
Pour water into the filler reservoir until its level stops falling. Fill the
reservoir to just beyond the half way point, stop, and repeat until the
level stops falling.
4.
Plug system in, turn on circuit breaker, and turn the laser key switch to
on.
5.
Make sure that the filler reservoir is still about half full. Add water if
necessary.
6.
Replace cap.
7.
Continue with laser start up. If a “water low” message is still present, call
Laserscope Customer Service at (800) 356-7600 from inside the US or +1
(408) 943-0636 from outside the US; or call your local Laserscope
distributor.
5.3 PREVENTATIVE MAINTENANCE SCHEDULE
For optimum performance of the GreenLight HPS
™
laser system, preventative
maintenance needs to be performed every six (6) months.