Technical Specifications
(*)
:
(*)
Specifications subject to change without notice.
(1) Individually compared against a NIST or PTB reference.
Warranty
All Laser Point instruments are warranted against defect in material and workmanship
for one year after invoice date. During the warranty period Laser Point will repair, or at
his sole option, replace at no charge components that prove to be defective, provided
the parts are returned, shipping prepaid, to Laser Point Customer Service.
This warranty does not apply if the instrument or any accessory has been damaged by
accident or misuse, or as a result of service or modification by other body than Laser
Point. No warranty is given by Laser Point on coatings and battery pack.
8
F
F
i
i
t
t
Instruction Manual
(Rev11- Nov 2008;
Firm. vers.11
)
1
Parameter
Fit 50 Fit 200 Fit 500
Maximum Measurable Power (W)
50 200 500
Minimum Measurable Power @ ±3% Accuracy (W) 1.5
(1)
5
(1)
15
(1)
Absolute Minimum Measurable Power (W)
1.25 3 7.5
Maximum Laser Beam Diameter (mm)
20 20 25
Power density damage threshold @ full scale at YAG
wavelength (W/cm
2
)
10000
Power density damage threshold @ full scale at CO
2
wavelength (W/cm
2
)
2500
Repeatability
±1%
ADC Resolution (W)
±0.06 ±0.25 ±0.60
Display Resolution (W)
0.01 0.1 0.1
Measurement Accuracy
±3%
Maximum allowable sensor head temperature (°C)
70
Time to measure and display data (s)
4
Power consumption in On status (mW)
26
Power consumption in Off status (
µ
W)
25
Power supply (2x AA Batteries) (V)
3
Continuous operation without battery replacement (h)
200
Operating temperature range (°C)
+10 to +40
Storage temperature range (°C)
-10 to +60
Weight (body) (g)
336
Weight (sensor with cable) (g)
178 200 280
Size (head) DxH (mm)
56x20 56x25 66x30
Size (body) LxWxH (mm)
95x71x46
WARNING! The user of this instrument must be trained to the
use of power lasers and associated risks. LaserPoint is in no way
liable for any damage resulting from improper, careless or above
rated limits use of this instrument.