I-Extract Instruction Manual
CEB00XXX-53-000
Page 8 of 11
Issue 2 30 January 2020
4.4 HEPA Filter Change
1)
Remove upper access door.
2)
Release the filter by dropping the lever arms of
the clamps as shown.
3)
Pull out the old filter from the filter section and
dispose of according to local regulations.
4)
Insert new filter taking care to respect any airflow
indication arrow. The filter seal should face
upwards to seal against the filter housing sealing
face.
5)
Clamp the new filter by pulling the lever arm of
the cambars into the vertical position.
6)
Refit access door and tighten hand knobs until
tight.
7)
Check door seal is good.