Horizontal Angle Compass Calibration Routine
Figure #19 shows the steps required to complete the Horizontal Angle Compass Calibration Routine.
To begin the routine, you should be holding the TruPulse 360
R and facing North. The instructions are
on the
Figure #19
At each step, wait approximately 1 second before pressing the
button.
Then wait another second before moving to the next position. It is important
that the unit is held steady when the
button is pressed.
The Horizontal Angle Compass Calibration routine can be aborted at any
time during the procedure with a long press of
or
. If the calibration
is aborted, the unit restores the previous stored calibration.