28
Actual case temperature
ATC
∗
-
°C
Error status
ERR
∗
-
-
Remark:
Some terminal programs send out the corresponding code right after the key was hit. Therefore a
wrong input must be finished with hitting the ENTER key on the PC and the input must be repeated.
Note:
The value for the temperatures, the upper and lower temperature limits must be within 0°C to 50°C.
Specific associations:
•
Laser operating mode (set LMO v)
Following associations of parameter values to be typed in are valid:
Value v
Laser operating mode
Remark
0
CW, NO TIME LIMIT
Default
1
CW, TIME LIMIT
2
PULSE, NO TIME LIMIT
3
PULSE, TIME LIMIT
•
Get condition laser (get LAS)
Following associations of values which will be returned are valid:
Value v
Condition laser diode
0
Laser diode off
1
Laser diode on
•
Set condition laser (set LAS)
The command for setting the condition (turning on or turning off) of the laser diode has a toggle
function. Each time the command set LAS is sent out the condition of the laser diode changes.
For example if the laser diode is presently turned off (as for example when the Laser Diode
Controller is just switched on with the Power switch), the laser diode will be turned on when the
command set LAS is sent out. On the other hand if the laser diode is presently turned on and the
command set LAS is sent out, the laser will be turned off.
Therefore we recommend to use the following programming structure:
First get the condition of the laser diode whether it is turned on or turned off with the command get
LAS. Depending on the returned value and the desired action by the user the decision to send the
command set LAS should be done.
•
Enable parameters (set AIE v)
The value for setting a “enable” parameter:
Value v
“enable” parameter
0
DISABLE
1
ENABLE
E.g. set AIE 1 enabled the analog input.