SoundAdvisor Model 831C
Setting Up a Measurement With Measurement History
15-5
Selecting Single Block Timer Mode With Measurement History
a.
Enter a
Start (y-m-d)
date and
Time
, and an
End (y-m-d)
date and
Time
, where
Time
is the time of day.
b.
Select
the
Enable Measurement History
checkbox.
c.
If you want to send a measurement interval exceedance alert, select
your preferred method in-line with Alerts, and select a Trigger Source
and Trigger Level (dB) for the alert. The Trigger Level cannot be less
than 0.1 (dB).
TAKE NOTE
Trigger Sources
for this alert include the following statistics:
LAeq, LCeq, LZeq, Lmax, Lmin
, and all 6 values of
Ln
. If the current dB level is
greater than the Trigger Level (dB)
of the selected Trigger Source
at any time
during the current measurement interval, the 831C sends an alert notification
via email or text when the interval ends.
d.
Select any additional settings on the Setup Manager tabs. For more
information, see
6.2.2 Customizing Setup Values On the Setup
e.
Press
Close
(
/*/
) and save your Setup.
f.
Set it as the Active setup file.
Selecting Daily Timer Mode With Measurement History
a.
Enter a
Start (y-m-d)
date and
Time
, and an
End (y-m-d)
date and
Time
, where
Time
is the time of day.
b.
If desired, select the checkbox in-line with
2
, and
3
to enter additional
measurement periods.
c.
Select
the
Enable Measurement History
checkbox.
d.
If you want to send a measurement interval exceedance alert, select
your preferred method in-line with
Alerts
, and select a
Trigger Source
and
Trigger Level (dB)
for the alert. The Trigger Level cannot be less
than 0.1 (dB).
TAKE NOTE
Trigger Sources
for this alert include the following statistics:
LAeq, LCeq, LZeq, Lmax, Lmin
, and all 6 values of
Ln
. If the current dB level is
greater than the Trigger Level (dB)
of the selected Trigger Source
at any time
during the current measurement interval, the 831C sends an alert notification
via email or text when the interval ends.
e.
Select any additional settings on the Setup Manager tabs. For more
information, see
6.2.2 Customizing Setup Values On the Setup
f.
Press
Close
(
/*/
) and save your Setup.
g.
Set it as the Active setup file.