  Caution

Always check the instrument rotation 

direction before using the device.

Motor LED Light ON/OFF: 

Press the LED light ON/OFF button (5) on 

the tablet to turn lighting on or off (Figure 

11).  

Airflow Method: 

Press the airflow button (6) on the tablet 

(figure 11) to toggle between the two 

modes.

• Progressive mode (one small arrow) 

allows the user to vary the handpiece 

speed, up to the maximum set rpm, using 

the foot pedal. 

• Instant mode (two small converging 

arrows) will instantly reach the maximum 

set rpm once the foot pedal is pressed. 

Voice Commands: 

Voice commands can be used to change 

speed presets, turn the handpiece LED 

lighting on/off, and change motor direction 

to forward/reverse.  (The motor must not 

be operating to use voice commands). 

• Be sure the microphone (7) symbol on 

the tablet display (figure 11) does not show 

a strikeout line indica�ng voice commands 

are disabled.  Touch the symbol once to 

enable voice commands. 

• The voice control wakeup command is 

“Hey Motor”.  The tablet will respond to 

“Hey Motor” with an audible beep 

indica�ng it is ready to respond to a 

command.  A visual �mer will also appear 

on the tablet screen indica�ng how much 

�me the user has le� to give the command 

before the tablet returns to sleep mode.

• To change a maximum rpm preset, simply 

say, “Hey Motor” followed by “One”, “Two”, 

or “Three” depending on the maximum 

rpm preset you want. 

• To change the motor rota�on direc�on, 

simply say, “Hey Motor” followed by either 

“Forward” or “Reverse”. 

• To turn the a�achment LED light on or 

off, simply say, “Hey Motor” followed by 

“Light On” or “Light Off”. 

  Caution

When using the voice activation, the end 

user must visually verify the desired 

selection is correct before utilizing system.

Steps For Normal Use: 

• 

Connect an a�achment to the motor. 

• 

Select the gear ra�o for the a�achment. 

• 

Select a maximum rpm preset (manually 

or by voice command) 

• 

Select the rota�on direc�on of forward 

(clockwise) or reverse (counter-clockwise) 

(manually or by voice command). 

• 

Select the air flow method of progressive 

or instant. 

• 

Press the dental unit foot pedal to start 

the Percep�ve motor. 

  Caution

Unplug power supply from mains to safely 

terminate operation of the Perceptive 

System. 

Front panel LED:

Meaning:

GREEN(steady)

Normal opera�on

RED (blinking)

Overcurrent

RED (steady)

Drive Fault

WHITE (pulsa�ng)

System Low Power 

Mode

General Settings:

Additional motor control settings may be 

adjusted by first pressing the Settings (8) 

button (Figure 11).  These consist of (Figure 

12):

-Display sleep:  adjusts time of non-use 

before tablet display goes into sleep mode. 

-Pressure:  select to display air pressure in 

PSI or BAR. 

-Variable Pressure Mode: FULL uses the 

entire air pressure range of the foot control 

to change motor speed.   LIMIT only uses a 

portion of the air pressure range. (Full is 

recommended.)

-Status LED:  Adjusts the brightness of the 

motor control box LED status indicator. (See 

below for description of Status LED signals). 

-Motor LED Brightness:   Adjusts the 

brightness of the motor LED output and 

therefore the attachment light output 

brightness. 

-Motor LED Fade-in:  Adjusts how fast the 

light comes on after foot control 

depression. 

-Motor LED Fade-out:  Adjusts how fast the 

light goes out after foot control is released.

-Motor LED Time-out:  Adjusts how long 

the light remains on (delayed shut off) after 

motor operation ends.

 

To return to the main screen, touch the 

back arrow at the top left of the screen.

Motor Control Box Status LED Meaning:

The front panel of the motor control box is 

equipped with a status LED light with the 

following meanings: 

Device Maintenance

  Caution

• Wearing of personal protective 

equipment (gloves, goggles etc.), should be 

complied with by medical personnel using 

or performing maintenance of medical 

devices that are contaminated or 

potentially contaminated. 

• Pointed and sharp instruments should be 

handled with great caution. 

• In the event of prolonged disuse, the 

instrument must be stored in a dry 

environment. 

• Products containing acetone, chlorine 

and bleaches are not recommended as 

disinfectants. To keep the surfaces of the 

hose in good condition, it is advisable to 

periodically wipe the complete length of it 

with a cloth dusted with talcum powder. 

Do not immerse in disinfectant solution. 

Do not immerse in ultrasonic bath. 

• We recommend that the motor control is 

cleaned as directed below before the initial 

first use and subsequently after each 

treatment. 

Cleaning-disinfection Of External Surfaces:

 

Clean and disinfect the external surfaces of 

the Perceptive motor control unit, tablet, 

and hose by gently rubbing for about 15 

seconds with a clean cloth soaked in 

Isopropyl alcohol. 

General Precau�ons

 USA Caution: Federal law restricts this 

device to sale by or on the order of a 

Dentist.

Other Precautions for Use:

• The device must be used by a qualified 

person in accordance with the current legal 

provisions concerning industrial safety, 

health and accident prevention measures, 

and these working instructions. 

In accordance with these requirements, 

the operators: 

Must only use operating devices that are 

in perfect working order. In the event of 

irregular functioning, excessive vibration, 

abnormal heating or other signs indicating 

malfunction of the device, the work must 

be stopped immediately; in this case, 

contact Lares Research for repair. 

-Must ensure that the device is used only 

for the purpose for which it is intended, 

must protect themselves, their patients 

and third parties from any danger, and 

must avoid contamination through the use 

of the product.

-Rest the device on a suitable support to 

avoid risks of infection for yourself, the 

patient or third parties. 

-Excess material from products used for 

maintenance (lubricants, cleaning products 

and disinfectants) originating from the 

attachments may penetrate into the 

electric motor and interfere with its 

functioning. It is essential to follow the 

maintenance instructions accompanying 

each product.  Never lubricate the electric 

motor. 

Recommendations:

• It is essential to use dry, purified 

compressed air to ensure the long working 

life of the device. Maintain the quality of 

the air and water by regular maintenance 

of the compressor and filtration systems. 

The use of unfiltered hard water will lead 

to early blockage of the tubes, connectors, 

and spray ports. 

• The device must not be used in the 

presence of open lesions, injury to soft 

tissue or recent extractions. The exhaust air 

could propel infected material into the 

wounds and cause infections and risk 

embolism. 

• The device is intended for medical 

treatment only; any use other than that for 

which this product is intended is 

unauthorized and may be dangerous. The 

medical device meets all the current legal 

requirements.

 

Warranty and Service

Warranty:

Each Percep�ve motor control box, power 

supply and supply hose is warranted 

against defects in materials and 

workmanship for a period of 2 years from 

the date of purchase.

Addi�onal Condi�ons of Warranty:

1. Warranty registra�on is automa�c as of 

shipping date (Outside the US warranty 

registra�on may be required). 

2. The product must be operated and 

maintained in accordance with procedures 

outlined in these instruc�ons. 

3. The product must not be subjected to 

abuse and/or neglect. 

4. The product must not be repaired or 

disassembled by anyone other than Lares 

Research or your authorized Lares 

distributor. 

Lares Research will repair or replace at its 

discre�on without charge, any defec�ve 

parts covered by this warranty provided 

the Percep�ve motor control box, power 

supply and supply hose is returned to the 

factory, transporta�on prepaid. (Outside 

the US return to your authorized Lares 

distributor.) Lares Research makes no other 

warran�es expressed or implied. 

Servicing: 

Never disassemble the device. For any 

modifica�on or repair, we recommend that 

you contact your Lares account manager 

directly. Lares Research recommends that 

you have the device checked or inspected 

once every 3 years.

Transporta�on, Storage & Disposal

Transport and Storage Condi�ons: 

Temperature between -25˚C (-13˚F) and 

70˚C (158˚F), rela�ve humidity between 

20% and 80%, atmospheric pressure 500 

hPa to 1060 hPa (7.25 to 15.37 psi). 

Do not store the motor with a contra angle 

a�ached for extended periods of �me. 

Disposal: 

This device must be recycled. Electrical and 

electronic equipment may contain 

dangerous substances which cons�tute 

health and environmental hazards. The 

user must return the device to an approved 

body for treatment and recovery of this 

type of equipment (European Direc�ve 

2002/96/EC). 

Troubleshoo�ng 

Motor Error:

When an error is detected on one of the 

motor phases, an error tone (three long 

beeps) will sound once on the tablet, and 

the icon above will keep blinking as long as 

the drive is ac�vated by air pressure. To 

reset the error, release the pressure to 

deac�vate the drive and ensure 

connec�vity of the motor. 

Temperature Warning:

A motor overheat protec�on system is in 

place. This will trigger: 

a) If average current over a long period of 

time exceeds 600mA, motor torque will be 

slowly limited until the average current falls 

below 600mA. 

b) If peak current exceeds 2500mA, motor 

torque will be quickly limited until peak 

current drops below 2500mA. 

This issue is commonly caused by a faulty 

attachment (contra-angle, etc). If the issue 

ony exists with the attachment on the 

motor, it is likely time to have that 

attachment repaired. Contact Lares 

Research with any questions. 

When any of the above conditions occur, 

the icon above will blink on the tablet, and 

a warning signal (three short beeps) will 

sound. The warning signal will keep 

repeating until the unit has exited the 

overheat condition. 