CLIENTS user guide SA1000
Page 5
2.0 LaplaCell 600 Specification
2.1 Performance
General
Usable EUT volume:
600mm cube
EUT weight:
100kg (Spread load)
Internal size:
780mm x 870mm x 820mm
Calibrated frequency range: 30MHz - 2GHz (in two ranges)
Screening attenuation:
<60dB
Internal damping:
Ferrite absorbers
Immunity mode
Max field strength with
10W power amplifier:
10V/m minimum
Field uniformity:
6dB within EUT volume measured as recommended by IEC
Power/field ratio:
> 1.0V/m.per dB
Max input power:
50W
Sensor output level:
0 - 2.6V corresponding to internal field strength.
Calibration data:
Supplied with each cell.
Sensor supply
12v dc input @800mA
Power input Impedance:
50ohm
Input VSWR:
1.5 : 1
Input connector:
N Type
Emissions mode.
Output Impedance:
50ohm
Output connector:
N type
I/O facilities
Integrated filter panel for power and signal connections into the cell.
Basic unit offers....
.mains input,
10way general I/O
plus illumination and camera socket.
Additional interconnections may be specified at time of ordering.
Monitoring
Optionals:
Internal lighting
Internal video camera
Forced ventilation
Additional EUT connections
Interlocks:
Door fitted with safety switch.
Physical
Door aperture:
520mm x 840mm
Weight:
400kg
Size:
length
3.15m
width
1.32m
height
1.57m
Outer construction:
Stainless steel
Internal EUT chamber:
Polypropylene.
Septum:
Double tapered wire guided.