Safety Instructions 

 

Observe the operating and safety instructions for all devices used in the set-up. 

 

 
 
Never use any damaged or defective devices. 

 
 
 

 

The  use  of  this  near-field  probe  to  test  devices  with  voltages  higher  than  safety 
extra-low  voltage  (SELV)  is  only  permitted  for  trained  personnel.  Additional 
protective measures such as using an isolating transformer, additional insulation, 
etc.  must  be  implemented  to  safeguard  against  electric  shocks  and  to  minimize 
the risk of injuries. 
 
Depending  on  the  measured  field  strength,  voltages  exceeding  the  permissible 
maximum input voltage of the connected device may be generated at the probe's 
output.  Under  certain circumstances,  this  may  also occur  with short-term  events, 
such  as  short-circuits,  high  inrush  currents,  and  so  on.  Employ  protective 
measures,  such  as  using  attenuation  elements,  transient  limiters  or  similar 
devices, as required. 
 
The probe tip is sensitive to mechanical stress

 

and is not a guaranteed insulation. 

The  user  is  therefore  required  to  provide  appropriate  insulation  to  protect  the 
device under test and the connected measuring device. 

 

Technical Parameters 

 

Upper frequency limit 

40 GHz 

Lower frequency limit 

Depending on the measuring device 

Resolution 

0.2 mm 

Inner distance of the measuring electrode to 
the measuring tip 

about 0.5 mm 

Output resistance 

5

0 Ω 

Connection - output 

2.92 mm (K), female, jack 

Weight 

19 g 

Sizes (L x W x H) 

(145 x 9 x 9) mm 

Table 1: HR-E 40-1 technical parameters

 