LANDTEC GEOTECH DIVEAIR 2 Operating Manual Download Page 9

Page 9  

Geotechnical Instruments 

CODAOM002 Iss.06

 

3.5  Internal VOC filter 

VOCs can affect the reading from the CO sensor. To get over this problem the Diveair analyser has a 
filter to remove the VOCs before the gas reaches the CO sensor. The filter is situated within the monitor 
and should not normally need replacing between services. 
 
The lifetime of the filter is dependent on the amount of VOCs that it is required to remove.  If 
contamination of the filter is suspected it can be examined and changed by opening the case. The case 
should be opened only in a clean and dry environment. Replacement filters can be obtained from 
Geotechnical Instruments. 

4

Servicing and Maintenance 

It is recommended that instruments are serviced regularly by the manufacturer. 
There are no user serviceable components within the analyser apart from those mentioned below. 
 

4.1  Particulate Filter Change 

The analyser contains a particulate filter located within the inlet port of the instrument. The filter is 
changed as follows:– 

1.  Remove the sampling tube from the analyser by carefully removing the inlet nozzle. 
2.  The filter inlet nozzle is removed by using an Allen key (Hex key) to unscrew the inlet nozzle. 

When the nozzle is removed, the old filter will be attached to the rear of the nozzle. This filter can 
be removed by hand. 

3.  Carefully locate a new filter onto the rear of the nozzle, ensuring it is firmly located. Carefully 

reposition the nozzle, ensuring the sealing O-ring is in place in the nozzle housing and tighten 
using the Allen key (Hex key). 

 

4.2  Water/PTFE Filter Change 

The Water/PTFE filter consists of a small plastic housing containing a PTFE membrane that has push on 
tube fitting at one end and a luer fitting the other. 

The analyser should be opened only in a clean and dry environment. 

To change the Water/PTFE filter:- 

1.  Switch the analyser off. 
2.  Open the analyser. This may require the removal of the security tags (cable ties) before the 

analyser can be opened. 

3.  Carefully twist the leur fitting to undo the connection and release it from the holding clamp. 
4.  Remove the tube from the other end of the filter. 
5.  Take the new filter and push the tube onto the hose barb connector on the filter. 
6.  Put the filter through the clamp and attach and tighten the leur fitting. 
7.  Close the analyser and replace the security tags. 
8.  Dispose of the old filter. 

 

4.3  VOC Filter Change 

The VOC filter consists of a cartridge that has push on tube fittings at either end. 

The analyser should be opened only in a clean and dry environment. 

To change the filter:- 

1.  Switch the analyser off. 
2.  Open the analyser. This may required the removal of the security tags (cable ties) before the 

analyser can be opened. 

3.  Carefully remove the plastic tube from both ends of the filter tube. 
4.  Insert the new filter and push the tubes onto the hose barb connectors on the filter. 
5.  Close the analyser and replace the security tags. 
6.  Dispose of the old cartridge. 

 
If the contents of the filter escape: 

No special disposal procedures required 

 

Wash hands after handling 

 

Consult medical advice if ingested 

 

Summary of Contents for GEOTECH DIVEAIR 2

Page 1: ...air analyser OPERATING MANUAL Geotechnical Instruments UK Ltd Sovereign House Queensway Leamington Spa Warwickshire CV31 3JR England Tel 44 0 1926 338111 Fax 44 0 1926 338110 Email sales geotech co uk www geotech co uk ...

Page 2: ... settings 7 3 3 Alarm Settings 8 3 3 1 Setting alarm limits 8 3 3 2 Setting alarm mode 8 3 4 Storing and reviewing readings 8 3 4 1 Logging Options 8 3 4 2 View data 8 3 4 3 Clear Memory 8 3 5 Internal VOC filter 9 4 SERVICING AND MAINTENANCE 9 4 1 Particulate Filter Change 9 4 2 Water PTFE Filter Change 9 4 3 VOC Filter Change 9 4 4 Replacement of the Oxygen Cell 10 4 5 Battery Replacement 10 4 6...

Page 3: ...composition of compressed air used to fill gas cylinders for underwater diving operations As well as being able to measure the concentrations of Carbon Dioxide and Oxygen the analyser can also measure possible contamination due to Carbon Monoxide and Volatile Organic Compounds VOCs The Diveair 2 Analyser will measure Carbon Dioxide Oxygen Carbon Monoxide and VOCs Main Features include Portable wit...

Page 4: ... on the analyser Failure to do so could result in dirt entering the analyser which could cause damage to the optics When the Gas Analyser is in use ensure the exhaust port is clear and unblocked at all times If the exhaust port is blocked while the Gas Analyser pump is running or the analyser is being checked for accuracy with a regulated gas supply the analyser could become over pressurised Such ...

Page 5: ...ing key 9 Batt from the main gas reading screen press any key to exit It is recommend that you wait at least five minutes after powering on before checking this graph to give the battery output time to stabilise after powering up If the analyser is used infrequently the battery capacity may be reduced To restore the battery to full capacity the unit should be totally discharged and then charged fu...

Page 6: ... span is checked It is recommended that a zero and span is performed on first receipt of the instrument On receipt of the instrument or when the instrument has not been used for some time a zero and span calibration should not be carried out until the VOC sensor has normalised The instrument should be on for a minimum of 1hour before calibration 3 2 1 Zeroing The Diveair analyser can be zeroed by ...

Page 7: ...channel either okay or FAILED There are a number of reasons for a particular channel to fail to span calibration see troubleshooting 8 Turn off the gas and disconnect the hoses 9 Switch the pump on and purge the analyser with ambient air for 5 minutes 3 2 4 Trouble Shooting Zero calibration failed A possible reason for this is because the instrument is trying to zero to a level which is outside th...

Page 8: ...nt that this is set correctly as every reading logged includes the time and date at which it was recorded To do this press key 1 Check time date and follow the simple on screen instructions The logging interval time can also be set between 1 and 60 minutes From the logging menu press key 2 Set interval and follow the instructions The final option on this menu allows the user to start and stop data...

Page 9: ...e ensuring the sealing O ring is in place in the nozzle housing and tighten using the Allen key Hex key 4 2 Water PTFE Filter Change The Water PTFE filter consists of a small plastic housing containing a PTFE membrane that has push on tube fitting at one end and a luer fitting the other The analyser should be opened only in a clean and dry environment To change the Water PTFE filter 1 Switch the a...

Page 10: ... is switched off 2 Open the analyser This may require the removal of the security tags before the analyser can be opened The analyser should only be opened in a clean and dry environment 3 Locate the flying lead from the battery pack Unplug it from the printed circuit board 4 Undo the two screws holding the battery pack in place The battery can now be removed from its holder 5 Slide the new batter...

Page 11: ...ight 5 4 Ports Inlet Tube Nozzle with internal fibre filter Outlet Tube nozzle Electrical 7 Pin Lemo sealed communications connector Power 6 Pin Lemo charger socket and external power socket 5 5 Filters Inlet 0 01µm FIBRE filter operator changeable Internal 0 2µm PTFE hydrophobic filter not field changeable Geotechnical Instrument VOC filter not field changeable 5 6 Carbon Dioxide Sensor Type Infr...

Page 12: ...surement and temperature ranges within 8 hrs of calibration T90 Response Time 20 seconds Cross Gas Effects Unaffected by CO2 Drift 5 signal loss year Pressure Compensation Diffusion sensor atmospheric 10 Anticipated Life Two years in air Maximum Overload 1 500ppm Humidity Range 10 90 non condensing Calibration User zero span re calibration facility 5 9 VOC Detector Type Photoionisation detector Ra...

Page 13: ... 9 23 Butanethiol 1 9 14 Nitrile 3 10 39 Butene 1 9 58 Butene cis 2 9 13 Butene trans 2 9 13 Butyl acetate n 10 Butyl acetate sec 9 9 Butyl alcohol n 10 06 Butyl alcohol sec 9 88 Butyl benzene n 8 69 Butyl benzene s 8 68 Butyl benzene t 8 68 Butyl formate n 10 5 Chemical IP Butyl mercaptan 9 15 Butylamine 8 71 Butylamine s 8 7 Butylamine t 8 64 Butyltoluene tert 8 28 Butyne 1 10 18 Butyraldehyde n...

Page 14: ... iodide 10 38 Hydrogen selenide 9 88 Hydrogen sulfide 10 46 Hydrogen telluride 9 14 Hydroquinone 7 95 Chemical IP Iodine 9 28 Iodo 1 2 methylpropane 9 18 Iodobenzene 8 73 Iodobutane 1 9 21 Iodobutane 2 9 09 Iodopentane 1 9 19 Iodopropane 1 9 26 Iodopropane 2 9 17 Iodotoluene m 8 61 Iodotoluene o 8 62 Iodotoluene p 8 5 Isobutane 10 57 Isobutyl acetate 9 97 Isobutyl alcohol 10 12 Isobutyl formate 10...

Page 15: ...0 Picoline 2 9 02 Picoline 3 9 02 Picoline 4 9 04 Propanethiol 1 9 2 Propargyl alcohol 10 51 Propiolactone 9 7 Propionaldehyde 9 98 Propionic acid 10 24 Propyl acetate 10 04 Propyl alcohol 10 2 Propyl benzene 8 72 Propyl ether 9 27 Propyl formate 10 54 Propylamine 8 78 Propylene 9 73 Propylene imine 9 Propyne 10 36 Proylene oxide 10 22 Pyridine 9 32 Pyrrole 8 2 Quinone 10 04 Stibine 9 51 Styrene 8...

Page 16: ...s producers responsible from July 1 st 2007 in meeting their obligations with the fundamental aim of reducing the environmental impact of electrical and electronic equipment at the end of its life Geotechnical is now registered with the Environmental Agency as a producer and has joined a recycling scheme provider who will manage and report on our electrical waste on our behalf Our Producer Registr...

Reviews: