OMBG5KN4.7.1
BIOGAS 5000 Gas Analyzer
Page
20
of
61
Power supply front and back drawing:
Charger: Input voltage:
100-240V AC +/- 10%
Input frequency:
50-60Hz +/- 10%
Input current:
0.4A@100VAC .. 0.2A@240VAC
Output voltage:
10.1VDC max
Output current:
1.5A max
Note:
This charger has been internally restricted to 1.5A
6.10
Cleaning instructions
Do NOT use any cleaning agents to clean the analyzer or battery charger as they may
have an adverse effect on the safe use of these devices.
6.11
Memory
The analyzer's memory is stored in a readings and configuration database. The analyzer
will prompt when its memory is full, and you will not be able to store any further
readings. Please download your readings via LSGAM or the Basic Download Software and
then clear the memory.
Note: The analyzer should never be stored for prolonged periods with valuable data
in its memory. It is advisable to download all readings to LSGAM at the end
of each day’s monitoring. To clear the memory, please refer to the LSGAM
operating manual.
6.12
Warning and error codes
When switched on, the instrument will perform a predetermined self-test sequence
taking approximately ten seconds. During this time many of the instrument’s working
parameters and settings are checked. If any operational parameters are out of
specification or if the pre-programmed recommended calibration/service date has
passed, errors or warnings may be displayed.
Note: For further information please refer to section ‘10.0 Problem Solving’.