Landis+Gyr Ultracold T550 Technical Description Download Page 32

Communication 

 

32/49 

 

**) If the battery of the meter is of any other type, it must be replaced by a battery of 

type D. This battery life time is valid for the standard data telegrams (P600, P601) 
and T550 with standard measuring interval for flow and temperature. 

9.7 

Radio module 868 MHz LoRaWAN (CMi4110) 

 

Note:

 For T550 with firmware 

≥ 5.15.  

 

Note:

 

The module is displayed as “MB G4” on the LCD. For de-

tails see also Quick Manual and User Manual.  

 

Note:

 Can be ordered as an accessory with internal or external 

antenna.

 

 

The LoRaWAN module sends meter data via LoRaWAN network 

1)

 to an ap-

plikations Server. The end customer also has access to the application server. 
The meter data is sent immediately after the start. The module can be param-
eterized with the mobile APP 

“LoRaWAN MCM” from Elvaco.  

Frequency 

868 MHz  

Transmission power (ERP) 

25 mW (14 dBm) 

Range*) 

 

      free field 

10 km 

      inside buildings 

100 m 

Power supply 

 

      via calculator **) 

Battery type D 
Transmission interval 60 minutes bat-
tery for 11 years 
Transmission interval 160 minutes 
battery for 16 years 

      via power supply 

110/230/24 V (not with Elvaco 24V) 

Standard 

LoRaWAN Version 1.0 (

Class A bi-

directional)

 

 

Long Range Wide  Area Network

 (

LoRaWAN

) is a low-power  wireless network pro-

tocol. The LoRaWAN specification is defined by LoRa Alliance is freely available and 
uses a special modulation technique. 
*) 

Depending on the structure of the building can deviate significantly.

 

**) 

If the battery of the meter is of any other type, it must be replaced by a battery of type D. 

This battery life time is valid for the standard data telegrams (P600, P601) and T550 with 

standard measuring interval for flow and temperature. 

9.8 

GPRS module 

The GPRS module is used for data acquisition over a mobile network

*

), using 

open standard protocols

**

) in push mode (as Email, HTTP, FTP, SMS) or pull 

mode as transparent M-Bus (GSM, TCP). The integration into billing systems 
happens via selectable report templates.  

The module is configured by SMS messages. A firmware update is possible at 
any time (“Over the Air“).  

The module includes an integrated M-Bus Master, which allows to read out up 
to 8 additional M-Bus Meter (so-called slaves). The meter data can be stored 
and send at configurable 

intervals. In addition, “ad hoc“ readings are possible. 

Summary of Contents for Ultracold T550

Page 1: ...UH50 ULTRACOLD T550 UH50 Flow sensor T550 UH50 Technical description 32 16 101 001 h Date 23 08 2018 Landis Gyr GmbH Ultrasonic heating and cooling meter T550 UH50 Residential district heating local...

Page 2: ...ing to EN1434 2014 1 1 000 total range Power measurement with maximum values tariffs selectable Data logger for system monitoring 60 monthly values Logbook Battery or mains operated Optical interface...

Page 3: ..._____19 Service loop 2 LOOP 2 ___________________________________________________19 Service loop 3 LOOP 3 __________________________________________________20 Service loop 4 LOOP 4 ___________________...

Page 4: ...D ______________________________________ 35 11 Error messages__________________________________________________________37 12 Log functions ___________________________________________________________38...

Page 5: ...onsumption from the volume and temperature difference The meter combines modern microcomputer technol ogy with innovative ultrasonic measurement technology for which no mechani cally moving parts are...

Page 6: ...he electronic unit Be aware of sharp points on the thread flange and meas uring tube Only personnel trained in the installation and operation of meters in heating and cooling systems may install and r...

Page 7: ...ot dispose of the meter and the batteries with domestic waste Observe the local stipulations and laws on disposal You can return the lithium batteries to the manufacturer for appropriate disposal foll...

Page 8: ...eshold f T 0 2 K Temperature difference T 3 K 120 K Temperature measurement range 0 180 C LCD 7 digit Optical interface Standard EN 62056 21 Communication Optional e g M Bus Separability Always option...

Page 9: ...15 6 160 3 8 1 2 3 2 5 130 G1 5 25 10 200 5 6 1 8 1 5 2 5 190 G1 5 25 10 210 5 3 1 7 1 5 2 5 190 DN20 5 25 10 210 5 3 1 7 3 3 5 260 G 1 1 4 7 35 14 60 14 4 5 3 3 5 260 DN25 7 35 14 60 14 4 5 5 6 150...

Page 10: ...c m h bar mm mm mm UH50 x03 0 6 16 130 G1 58 UH50 x04 0 6 25 130 G1 58 UH50 x05 0 6 16 110 G 48 UH50 x06 0 6 25 110 G 48 UH50 x07 0 6 16 190 G1 88 UH50 x09 0 6 25 190 G1 88 UH50 x21 1 5 16 110 G 48 U...

Page 11: ...ers with thread Fig 5 Overview dimensions of large meters with thread Order No qp PN a b c m h bar mm mm UH50 x45 3 5 16 260 59 G 1 B UH50 x47 3 5 25 260 59 G 1 B UH50 x50 6 16 260 59 G 1 B UH50 x60 1...

Page 12: ...0 x65 15 25 50 270 59 155 125 18 4 102 20 UH50 x69 15 25 50 200 59 155 125 18 4 102 20 UH50 x70 25 25 65 300 52 185 145 18 8 122 22 UH50 x74 40 25 80 300 56 200 160 18 8 138 24 UH50 x82 60 16 100 360...

Page 13: ...circuit as the meter Observe the admixtures Seal the temperature sensors and the fittings to protect against manipula tion Loosen the elastic band or the cable tie provided for the transport from the...

Page 14: ...s for sensor adapter set temperature sensor directly immersed A mounting set is included for meters with 5 2 45 mm temperature sensors It allows for mounting the temperature sensors directly immersed...

Page 15: ...e wall Make a loop downwards in order to prevent condensation running along the connected lines into the electronic unit Transducer cover Fig 14 Recommended installation position for cooling metering...

Page 16: ...Dimension of electronic unit 16 49 5 Dimension of electronic unit Fig 15 Dimensions of electronic unit Fig 16 Plan view and cross section of adapter plate...

Page 17: ...button To call up the pa rameterization op eration of the me ter Accessible after removing the cover 2 Button 2 Switches to the next display value within a loop 3 Button 1 Switches to the next loop 4...

Page 18: ...h display range and data displayed can differ from the description depending on the appliance parameterization Certain button functions can also be blocked Display values LOOPs User loop Service loop...

Page 19: ...ent temperature cold side in 2 sec cycles Operating time Operating time with flow Missing time Property number 8 digit Date Yearly set day DD MM Energy previous year on set day Volume previous year on...

Page 20: ...on the set day in 2 sec cycles with date stamp Max power on the set day in 2 sec cycles with date stamp Max temperatures on the set day in 2 sec cycles with date stamp MV for hot side or MR for cold...

Page 21: ...the loop Current tariff in 2 sec cycles with threshold value 1 Measurement interval flow Measurement interval temperature Module 1 M Bus module M Bus primary address 1 M Bus secondary address 8 digit...

Page 22: ...ld side 6 3 Monthly values The meter stores the following values for 60 months on the monthly set day Energy meter status Volume meter status Tariff register meter status Missing time meter status Flo...

Page 23: ...Resolution of the display 23 49 7 Resolution of the display Note The number of places after the decimal point of a value is based on the chosen measurement path and the chosen dimension...

Page 24: ...lly whether it is being powered from a battery or power supply Power supply requirements Requirements for measuring timebase Q 4 s and time base T 30 s 6 years 11 years 16 years Power supply 230 110 V...

Page 25: ...oltage 85 121 V AC or 196 253 V AC Type Safety class II Frequency 50 60 Hz Line voltage fluctuations maximal 10 of the nom voltage Overvoltage category II per EN60010 2500 V pulse voltage Power consum...

Page 26: ...the following communication modules Pulse module CL module M Bus module G2 M Bus module G4 M Bus module G4 MI with 2 pulse inputs Analog module Radio module 868 MHz GPRS module These modules do not h...

Page 27: ...yes yes MB MI yes yes yes yes yes yes 1 no yes yes CL yes yes yes yes 1 yes 1 no yes yes yes Restrictions only 1 module with fast pulses is possible only permissible on slot 2 min pulse duration 2 ms...

Page 28: ...ermits the output of pulses that can be derived from the quantity of heat the volume tariff register 1 tariff register 2 or the mistake sta tus Two channels are available whose functions can be parame...

Page 29: ...description TKB 3436 9 3 M Bus module G4 The M Bus module enables the meter to communicate with an M Bus center in order to transmit measured values Standard EN 1434 3 EN 13757 2 3 Protocol EN 60870 5...

Page 30: ...e TKB3448 Pulse Number of pulse inputs 2 pulse inputs per MI module Life of the module battery 5 years of operation 1 year storage duration if the M bus voltage is applied for at least 90 of the opera...

Page 31: ...vailed C1 mode OMS 4 1 2 with security profile B The radio module 868 MHz enables the meter to communicate with a center receiver using 868 MHz radio frequency The module supports OMS1 compliant data...

Page 32: ...ttery for 16 years via power supply 110 230 24 V not with Elvaco 24V Standard LoRaWAN Version 1 0 Class A bi directional Long Range Wide Area Network LoRaWAN is a low power wireless network pro tocol...

Page 33: ...Indication green yellow red M Bus standard 13757 M Bus Baud rate 300 and 2400 Bit sec Transparent M Bus Listening server on TCP and GSM data Maximum connected M Bus slaves 8 Maximum cable length 1000...

Page 34: ...er Summation in the relevant tariff register is only performed if the relevant threshold is exceeded Threshold 1 exceeded Summation in tariff register 1 Thresholds 1 and 2 exceeded Summation in tariff...

Page 35: ...levant M bus command one of the 3 tariffs can be acti vated or all tariffs can be deactivated Surcharge quantity tariff by means of return temperature tariff T12 The quantity of energy is summated dep...

Page 36: ...2 sec cycles for T10 switching times in 2 sec cycles for T11 for T12 The contents of the tariff registers are displayed in the user loop after the quantity of energy For tariffs T2 T3 T4 T5 T6 T10 T11...

Page 37: ...e sensor Check cold side temperature sen sors replace if necessary F3 Electronics for temperature evaluation defective Exchange the meter F4 Problem with the power supply Battery flat Check connection...

Page 38: ...nth and for the past 18 months without time stamp Ser No Description 1 F0 Air in measuring tube 2 F1 Interruption temperature sensor hot side 3 F2 Interruption temperature sensor cold side 4 F3 Error...

Page 39: ...ons 39 49 20 Monthly set das parameterized 21 Master reset performed 22 All times deleted 23 Missing time deleted 24 Maxima deleted Note Read out is performed via the optical interface with the servic...

Page 40: ...uring period than 1 hour the largest value from the max imum values calculated within one hour applies Note Parameterization and read out are performed with the ser vice software Note Data transmissio...

Page 41: ...ogger Heating meter for mounting place hot side Operable as flow meter Cooling meter 6 12 C Combined heat cold meter Length of the control cable between measurement tube and electronic unit up to 5 m...

Page 42: ...flowrate 1 5 m h length 190 mm nominal pressure PN16 connection G 1 B 23 Nominal flowrate 1 5 m h length 190 mm nominal pressure PN25 connection flanged DN 20 24 Nominal flowrate 1 5 m h length 190 mm...

Page 43: ...Dial plate for Montenegro Serbian ME Dial plate for Macedonia Macedonian MK Dial plate for Mongolia Mongolian MN Dial plate for The Netherlands Dutch NL Dial plate for Norway Norwegian NO Dial plate f...

Page 44: ...to national regulations CL Compliant to MID class 2 M2 Compliant to MID class 3 M3 Compliant with CEN 1434 class 2 T2 Compliant with CEN 1434 class 3 T3 Compliant acc to national regulations TL 13 En...

Page 45: ...e information about lifetime of battery in chapter 8 Ordering example Order example OMS 2 9 1 P601 1 Protocol type 2 Sending interval 3 Encryption 4 Data telegram UH50 XYY0 Y 00 YXEX YYX 2 9 1 P601 E...

Page 46: ...e radio module retrieves data from the T550 calculator in a time grid of 15 minutes The transmission interval has a different parameterizable time grid The age of data indi cates the time difference b...

Page 47: ...1 200 5 6 E 3 5 260 G 1 1 4 DN25 55 15 G 6 150 G 1 1 4 190 14 F 6 260 G 1 1 4 DN25 140 16 H 10 200 300 G 2 DN40 130 28 I 10 300 G 2 110 30 J 15 270 DN50 110 45 K 15 200 DN50 95 49 L 25 300 DN65 105 7...

Page 48: ...Pressure loss 48 49 Alternatively the value can be taken from the diagram 250 10 100 1000 0 1 1 10 100 pressure loss in mbar flow rate in m h 21 02 2017 A BC DE F G H I J K L M N O P...

Page 49: ...49 49 Landis Gyr GmbH Humboldtstrasse 64 90459 Nuremberg Germany www landisgyr eu...

Reviews: