Cyclops 160B
User Guide
Page 27
8) Enter the
Window Compensation
>
On
sub-menu and amend the
Window Compensation factor value with the
and
keys until
the display temperature reads the same as the recorded value taken
from the known value source. You must now use this
Window
Compensation
value, as it is correct for the chosen viewing window/
port material (e.g. 0.880).
9) Select the
Continuous
temperature measurement option.
10) Aim the thermometer through the viewing window/port at the target
and focus the graticule onto the area under observation. Check that the
graticule is fully filled by the target, re-position if necessary
11) Press the trigger to start measurement. The main display and eyepiece
display are updated with the reading.
12) Release the trigger to stop measurement and freeze the last recorded
value on the main display.